9. Troubleshooting
Fault
Cause
Remedy
MEDUMAT Easy will not
switch on.
MEDUMAT Easy defective.
Arrange for repair.
Battery failing.
Replace battery in battery compartment
(10.1, page 57).
If ventilator still refuses to switch on,
have internal auxiliary battery replaced
by manufacturer or authorized
specialists.
Stenosis alarm
(excessive airway
resistance)
Airways obstructed.
Remove obstruction.
Kink or obstruction in ventilation
hose/mask/tube.
Remove kink or obstruction; if
necessary replace parts.
Tube incorrectly positioned.
Correct tube position.
MEDUMAT Easy defective.
Arrange for repair.
Disconnection alarm
(breathing system
interrupted).
Ventilation hose leaking/slipped
out.
Check connections.
Mask/tube incorrectly positioned.
Pressure gauge hose leaking/
slipped out.
MEDUMAT Easy defective.
Arrange for repair.
< 2.7 bar alarm (oxygen
pressure too low).
Oxygen cylinder nearly empty.
Change O
2
Oxygen valve closed.
Open oxygen valve.
Pressure reducer defective.
Replace pressure reducer.
Kink or blockage in oxygen hose. Take corrective action.
alarm.
Battery failing or fuse defective.
Replace battery in battery compartment
(10.1, page 57).
If ventilator still refuses to switch on,
have internal auxiliary battery replaced
by manufacturer or authorized
specialists.
+ –