Troubleshooting
55
EN
Visual alarms flashing, but
no acoustic alarm and no
audio response.
Short-term electronic problem or
defect in electronic system.
Switch off, then on again.
If error recurs, arrange for repairs.
Acoustic alarm, but no
visual alarm flashing.
Acoustic alarm sounds
and all visual alarms
flashing.
Alarm: Device
malfunction.
Device defective.
No audio response.
Audio response deactivated.
Activate audio response (6.10,
page 32).
MEDUMAT Easy is
functioning, but without
any displays.
Pressure gauge hose on
MEDUMAT Easy or on patient
valve slipped off.
Check pressure gauge hose.
Kink in pressure gauge hose.
MV too low.
Ventilation parameter(s)
incorrectly set.
Check ventilation parameter(s).
MEDUMAT Easy defective.
Arrange for repair.
Unusually high oxygen
consumption.
Leak in oxygen supply.
Find and eliminate leak (8.2, page 47).
MEDUMAT Easy will not
switch off.
Operating error.
Keep button pressed for at least 3
seconds.
Pressure gauge not
reading “0”.
MEDUMAT Easy defective.
Arrange for repair.
Fault
Cause
Remedy