Functional checks
53
EN
After two successive inspiration phases the
dis-
connection
alarm should be set off. If audio re-
sponse is enabled, the ventilator announces
“Check ventilation system and settings”.
Drop in O
2
pressure (<2.7 bar O
2
)
1. Open the valve of the oxygen cylinder
slowly
.
2. Switch on MEDUMAT Easy.
3. Close the valve on the oxygen cylinder. When
the oxygen pressure in the system has fallen be-
low 2.7 bar, the
<2.7 bar O
2
alarm should be set
off.
If audio response is enabled, the ventilator an-
nounces “Check pressure hose system and gas
supply”.
Power supply (
)
The alarm that indicates a failing battery is checked
automatically in the self test that runs when
MEDUMAT Easy is switched on.
The power supply is in order if no alarm is set off
when the valve on the oxygen cylinder is opened and
MEDUMAT Easy is switched on and starts to
function.
+ –