26
Using the ventilator
EN
the tidal volume set. Use suitable volume measuring
devices to check the tidal volume actually adminis-
tered.
Check respiratory parameters during ventilation.
If lung compliance diminishes during ventilation, the
ventilator will react by increasing the ventilation
pressure at constant volume.
6.5 Ventilation with PEEP valve
A PEEP valve can be fitted to the expiration tube on
the patient valve with an adapter.
This valve makes it possible to use positive end-expi-
ratory pressure (PEEP).
Please see the PEEP valve instructions for details of
settings.
6.6 Ventilation with filter
A conventional filter with standard 15/
22 connectors can be fitted on the inspiration tube
of the patient valve for hygiene purposes and to con-
dition the air for inspiration. This will increase both
Example of ventilation sequence before and after
lung compliance diminishes