16
Safety instructions
EN
•
Smoking and open flames are strictly prohibited
in the vicinity of all fittings containing or trans-
porting oxygen.
•
During assembly and when changing the oxygen
cylinder, only hand pressure should be used
when tightening the screw connections to the
cylinder and to the pressure reducer. Never use
tools for this purpose. Excessive tightening dam-
ages the screw threads and seals, and can cause
leaks.
•
Protect oxygen cylinders from accidental falls. If
a cylinder falls over, the pressure reducer or the
valve may break off and cause a violent explo-
sion.
Important
•
Always open the valve of the oxygen cylinder
slowly to prevent pressure damage to the other
fittings.
•
The oxygen cylinder should never be completely
emptied, as this may allow air containing mois-
ture to enter the cylinder and cause corrosion.
Ventilation/Operation
•
Both patient and ventilator must be kept under
constant observation during ventilation.
•
When connecting the patient valve, please ensu-
re that the direction of respiratory gas flow
(> patient >) is correct. Make sure that neither
the expiration tube nor the spontaneous breath-
ing tube on the patient valve are blocked or their
function impeded in any other way, e.g. by the
patient’s position.
Note:
•
Disposable hose system WM 28110 is intended
to be used only once.