Edges of the adhesive patch are peeling up.
Solution
Follow the next steps to apply the overpatch, if needed.
NOTE: The overpatch is only needed if you notice that the
edges of the adhesive patch are starting to come up.
1.
Fold the overpatch along the line that
divides the two halves of the liner, with
the liner side facing outwards.
Peel off the liner, starting in the centre,
where the two halves of the liner meet.
IMPORTANT
: Make sure the overpatch does not fold on itself
before or during application.
2.
Apply the overpatch so that it covers
the edges of the adhesive patch.
/ 16
Additional information
16.1
Customer Service information
For help with your GlucoMen Day CGM system, call Cu-
stomer Service. A list of Customer Service numbers can
be found in the
INFO
section of the app.
In case of emergency, contact your healthcare profes-
sional or emergency medical response.
16.2
Cleaning your system
The GlucoMen Day CGM System is for single patient use.
Clean in general
Your CGM system parts should be cleaned only as nee-
ded to remove visible dirt and debris. This does not re-
duce the risk for transmission of infectious diseases. For
more assistance, call Customer Service.
Clean the sensor insertion tool and transmitter
Clean the transmitter and exterior of the sensor insertion
tool with a moist (not wet) lint-free tissue. You should
use one of the following:
•
Water
•
Isopropyl alcohol 70% solution
•
Mild detergent
Carefully wipe dry all parts with a dry lint-free tissue.
CAUTION
•
Do not share the system parts with anyone, including
other family members! Do not use on multiple patients!
The sensor and sensor assembly are considered bioha-
zardous and can potentially transmit infectious diseases,
even after you have performed cleaning.
•
Do not clean the transmitter charger, as cleaning solu-
tions may cause USB port connection issues or corrosion
of electrical contacts.
• Do not clean the transmitter while it is in use or placed
in the charger.
• Be careful and gentle while cleaning the transmitter
contacts.