60
A
dditional informa
tion
/
English
CAUTION
•
Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could re-
sult in improper operation. If such use is necessary, this
equipment and the other equipment should be observed
to verify that they are operating normally.
•
Portable RF communications equipment (including pe-
ripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any
part of the GlucoMen Day CGM system, including cables
specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
16.7.1
Guidance and manufacturer’s declaration
— electromagnetic emissions
The GlucoMen Day CGM system is intended for use in
the electromagnetic environment as specified below. The
customer or the user of the GlucoMen Day CGM system
should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
16.7.2
Guidance and manufacturer’s declaration —
electromagnetic immunity
The GlucoMen Day CGM system is intended for use in the
electromagnetic environment specified below. The custo-
mer or the user of the GlucoMen Day CGM system should
assure that it is used in such an environment.
NOTE: U
T
is the a.c. mains voltage prior to application of
the test level.
NOTE: At 80 MHz and 800 MHz, the higher frequency
range applies.
NOTE: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
The GlucoMen Day CGM system complies with the
electromagnetic compatibility immunity requirements
defined in ETSI EN 301 489-1 v2.1.1, ETSI EN 301 489-
17 v3.1.1, and IEC 60601-1-2.
Emissions from nearby electronic equipment are not
likely to interfere with the GlucoMen Day CGM system
when used in a home healthcare environment. These
may interfere with the proper operation of the GlucoMen
Day CGM system. Moving away from or turning off these
electronic devices may allow communication.
If interference does occur and the following device
behaviour is observed:
•
Loss or incorrect CGM measurements
•
Loss of transmitter charging capabilities
The user can mitigate the situation by following the so-
lutions recommended in
CHAPTER 15
Troubleshooting.
Summary of Contents for A.MENARINI diagnostics GlucoMen Day CGM
Page 1: ...User Guide...
Page 65: ......