Symptom

Cause

Cure

Lower display shows 

MA

4-20 mA loop open

Check wiring. See sections 
4.4.4 and 8.1

Lower display shows 

FAIL

Major system fault

Refer to section 10.1

Readings are erratic

(1)

 Bubbles in solution

(2)

 Debris in flow

(3)

 Bad or sticky check valves

(1)

 See above

(2)

 Install T strainer at inlet

(3)

  Flush check valves as 

described in section 10.4 
or replace the valves

Readings are lower than 
expected

(1)

  Condensate or leaky 

measurement cuvette

(2)

 Measurement cuvette dirty

(3)

 Reagents bad or expired

(4)

  Buffer reagent not being 

dispensed

(1)

  Install desiccant cartridge 

kit

(2)

 Replace or clean cuvette

(3)

 Replace reagents

(4)

  Check buffer lines and 

check valves

Upper display flashes

Sample Over-Range

Check sample. Sample may 
be too high to read

Upper display shows  
nOnE while attempting  
to calibrate

No current reading displayed

Wait for CLX-Ex2 to post a 
reading

10.5 Diagnostic Chart

10.6 Technical and Customer Assistance

If for any reason assistance is needed regarding this instrument please 

do not hesitate to contact either the HF scientific Technical Service 

Department or the HF scientific Customer Service Department:

HF scientific

16260 Airport Park Drive, Suite 140

Fort Myers, FL 33913 

Phone: 239-337-2116

Toll Free: 888-203-7248

Fax: 239-454-0694

Email: [email protected] 

Website: www.hfscientific.com

11.1 Maintenance Schedule

The recommended schedule is shown below. It is important to replace 

the reagents to get reliable accurate readings from the CLX-Ex2. 

The Buffer supplied can last for up to one year. Each mixed indicator 

reagent will last for up to 90 days* when kept special cooling chamber. 

The one year kit includes one buffer bottle and four sets of indicator 

reagents. The Buffer reagent is mounted in a clip and is further held 

with a Velcro strap. The indicator reagent is in a 125ml bottle mounts 

inside an insulator and also mounts in a clip with a Velcro security 

strap. See section 11.2 for more information on the reagents.
The CLX-Ex2 is shipped with one CLX-Ex2 Tubing/Cuvette kit, HF part 

#09950EX. The kit consists of the following:

Qty

Part

1

Spare Buffer Cap Assembly

1

Spare Indicator Cap Assembly

8

Pump Tubes

1

Spare Cuvette

Two replacement sets of reagent cap/tubing assemblies are supplied 

with the CLX-Ex2; one for the buffer and one for the indicator.
Additional kits can be ordered from your local HF scientific distributor or 

representative. It is recommended to keep one kit on hand at all times.
Generally, both pump tubes should be replaced annually. Spare cap 

assemblies for both the buffer and the indicator are supplied and can 

be replaced when needed. 

Every 90 days

1.  The indicator reagent required for operating this instrument must be 

changed at least every 90 days. The buffer reagent can be used up 

to one year.

2.  The internal strainer should be checked and cleaned if necessary.
3.  The glass cuvette should be inspected. Check for excessive debris 

on the inside surface of the glass. It is suggested to keep a spare 

cuvette to replace when required. The old cuvette may be cleaned, if 

possible, for future replacement.

Flushing the System

Press the SERVICE button to stop the water flow. Remove old reagents 

and discard. Place the suction tubes in a small container of clean 

water. Press SERVICE to return to operation mode, press PRIME and 

then 

  to flush the system with water. Remove the suction tubes from 

the water Press PRIME and then 

  to remove most of the water.

NOTICE

After a PRIME the CLX-Ex2 will perform a water calibration (WCAL). 

It will take a few minutes to complete this procedure.

*The 1 year buffer and 90 day indicator reagent life is primarily used on 

ballast water applications. The assumption is based on the factory default 

of a 60 second cycle and up to 60 hours total operation per month.

Only access when the atmosphere has been tested and known to 

be safe. 

WARNING

!

11.0 Routine Maintenance

Please note that to perform these maintenance 

operations the cabinet door must be open and the 

instrument in maintenance bypass mode. See section 

5.3 for more information on this.

Specific condition of use: Parts of the enclosure are non-conducting 

and exceed the maximum permissible resistance according to the IEC 

60079-0. Therefore, to avoid electrostatic charge build-up, it must not 

be rubbed or cleaned with solvents or a dry cloth when installed/used 

within a potentially explosive atmosphere.

NOTICE

Summary of Contents for HF scientific CLX-Ex2 Series

Page 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Page 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Page 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Page 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Page 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Page 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Page 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Page 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Page 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Page 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Page 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Page 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Page 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Page 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Page 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Page 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Page 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Page 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Page 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Page 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Page 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Page 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Reviews: