background image

9

4.4.5 RS-485/4-20 mA cable Ferrite

To meet IEC requirements for RF radiated immunity a clamp-on type 

ferrite is supplied in the accessory kit. It should be placed on the RS-

485 or 4-20 mA cable outside, as close as possible to the CLX-Ex2. If 

both outputs will be used, an additional ferrite will be needed and can 

be ordered from HF scientific Catalog number 24560T.

4.4.6 Modbus Control

In addition to the above mentioned Remote Standby Connections, 

the same control can be achieved using Modbus control. Two special 

Modbus addresses are provided.
The reading cycle is initiated using the coil 00005 with default at False 

(0). Setting this address to True (1) initiates the reading cycle.
At any time the standby status can be check at Modbus Coil Address 

00005 or Modbus Input Address 10005. False (0) indicates standby 

and True (1) indicates the measurement cycle is taking place. Control is 

only available from the Coil Address 00005.

4.5 Installing Reagents

The CLX-Ex2 will require that two reagents be installed 

prior to operation. These are a buffer and an indicator. For 

reagent preparation refer to section 11.2 

Replacing and 

Installing the Reagents.

5.0 Air Sentinel Controller Operation

The Air Sentinel controller is mounted on the side of the main enclosure. 

This controller will ensure proper safe operation of the CLX-Ex2. If of all the 

required conditions are met, the controller will act as a power governor. It 

does this by monitoring the pressurization of the main enclosure with clean 

dry air. Should this air supply pressure fail the 

PRESSURE

 lamp on the  

Air Sentinel unit will turn RED and power to the rest of the instrument  

is disconnected.

5.1 Start Up

When power is applied to the CLX-Ex2 for the first time, 

ensure that the air supply is connected as stated in section 

4.3 and the cabinet is closed and locked. Once power is 

applied, the Air Sentinel will go through a purge period of 

5 minutes where at least five volumes of the enclosure are 

pushed through the CLX-Ex2 and out the Air Sentinel vent. During the 

purge period the 

TIMER

 lamp Air Sentinel will be RED indicating the 

purge is in process and no power is applied to the rest of the CLX-Ex2.
At the complete of the purge, the 

TIMER

 lamp will go GREEN and 

power will be applied to the rest of the system. As long as the proper 

air supply conditions are met the CLX-Ex2 will continue to operate.

Do not apply power to the Air Sentinel controller unless the area 

has been properly tested and is known not to contain explosive 

materials.

5.2 Indicator Lamps

The indicator lamps are a very quick way to determine to operation of 

the controller and thus the safety condition of the CLX-Ex2.

Lamp Color Matrix

Function

Default

All lamps GREEN

Safe Operation

Power ON

TIMER

 lamp RED 

Instrument purging or Bypass 
operation

Power OFF

PRESSURE

 lamp RED

Unsafe Operation

Power OFF

BYPASS

 lamp RED

Bypass Key Operating

Power ON

No Lamps on

No power applied to CLX-Ex2

No power applied  

everything off

Using the table above, you can quickly determine the operating 

condition of the instrument. Under normal operation all GREEN lamps 

should appear.

WARNING

!

Summary of Contents for HF scientific CLX-Ex2 Series

Page 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Page 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Page 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Page 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Page 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Page 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Page 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Page 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Page 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Page 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Page 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Page 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Page 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Page 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Page 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Page 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Page 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Page 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Page 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Page 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Page 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Page 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Reviews: