background image

15

8.9 LCD Backlight Brightness

The LCD backlight brightness may need to be adjusted. This is of 

particular interest if multiple instruments are located in the same area 

and it is desired for the entire group to have the same appearance. Ten 

levels are available. The factory setting brightness is 8.
Change the brightness by pressing the 

 or 

 button. When the 

desired brightness has been selected, press the 

  button.

8.10 RS-485 Parameters

These menus will only appear if the RS-485 is enabled (see 8.3). The 

factory setting is 8 Bit, no (nOnE) Parity, 1 Stop Bit.
Make selections using the 

 and 

 buttons then press the 

  button 

to move to the next menu.

8.11 Cycle Time

The cycle time can be changed using this menu. 

Please note that 

changing this menu will directly affect the volume of reagent 

that will be consumed.

Make selections using the 

 and 

 buttons then press the 

  button. 

Allowable setting is from 60 to 600 seconds (10 minutes).

NOTICE

The indicator reagent has a defined life after being mixed regardless 

of the cycle time setting. If the reagent is kept in the powered cooling 

chamber the life is 90 days.

8.12 Water Conservation

To conserve water, the flush time can be adjusted use as little water  

as possible.

NOTICE

This feature is generally not used in ballast water applications and 

should be set to OFF.

The instrument requires 110 seconds to complete its normal 

operations, when the 

WCON

 is turned 

On

, at the factory cycle time the 

instrument sits idle for about 40 seconds. This results in about a 25% 

water savings. The actual amount of water conservation is dependent 

on the incoming water pressure and the cycle time setting.

The use of this option may result in some slight loss of accuracy. The 

factory setting for this option is 

OFF

.

Make selections using the 

 and 

 buttons then press the 

  button 

to exit to 

AUTO

 mode and save all configuration settings. If the 4-20 

mA in section 7.1 is turned ON, there are two additional menus that will 

appear before returning to 

AUTO

 mode.

In addition to averaging, the CLX-Ex2 has a software filter that limits 

the change between consecutive readings to 20%. For example, 

the reading following a reading of 1.00 ppm could not be displayed 

as higher than 1.20 ppm or lower than 0.80 ppm. This filter also 

helps smooth out large changes and eliminate reading spikes. After 

3 consecutive readings, it is assumed that the large change in the 

readings is “real”, and the filter will be disabled. For example, the 

change between reading 1 and 2, and readings 2 and 3 will be limited 

to 20%, but the change between readings 3 and 4 will not be limited.

Note: This software filter is completely disabled when the averaging 

(AVG) is set to 1.

Summary of Contents for HF scientific CLX-Ex2 Series

Page 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Page 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Page 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Page 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Page 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Page 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Page 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Page 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Page 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Page 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Page 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Page 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Page 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Page 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Page 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Page 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Page 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Page 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Page 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Page 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Page 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Page 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Reviews: