background image

16

8.13 4mA Adjustment

If the 4-20 mA setting is turned ON (8.1 

Setting the 4-20 

mA Output)

, the following two menus will appear. The first 

menu outputs a constant 4 mA while allowing for a small 

amount of adjustment. The adjustment can be made using 

the 

 and 

 buttons. This adjustment will allow the operator to make 

the CLX-Ex2 agree with a PLC or SCADA system. The adjustment 

limits are ± 200 counts or about ± 0.2 mA.

This setting will be slightly different on each instrument as each 

CLX-Ex2 will be factory set to 4.00mA. Press the 

  button when 

adjustments are complete to save this setting and move on to the 

20mA adjustment.

8.14 20mA Adjustment

This menu operates similar to the previous menu. This menu outputs a 

constant 20 mA while allowing for a small amount of adjustment. The 

adjustment can be made using the 

 and 

 buttons. The adjustment 

limits are ± 1000 counts or about ± 1 mA.
This setting will be slightly different on each instrument as each CLX-Ex2 

will be factory set to 20.00mA.

When complete with the 20mA adjustment, press the 

  button to exit 

to AUTO mode and save all configuration settings.

9.0  Additional Features and  

Options

9.1 RS-485 Outputs

The CLX-Ex2 has the capability to operate in two different RS-485 

modes. Included are a simple communication mode and Modbus 

communications. Both modes will automatically configure and do not 

require any changes or selections.

9.1.1 Simple Communication

The CLX-Ex2 can provide basic communications over simple programs 

such as the Hilgraeve HyperTerminal that is included with most 

Microsoft Windows packages. The user could also use Visual Basic 

or other programs. The factory setting communication 

parameters are 8 bits, no parity and 1 stop bit. These can 

be changed in the Extended CONFIG menus 8.9 

RS-485 

Parameters.

The master computer will send out:

•  Byte #1 the attention character “

:

” in ASCII or 

3A

 Hex

•  Byte #2 the address of the CLX-Ex2 being queried
•  Byte #3 & 4 CR LF or 0D 0A in hex

The CLX-Ex2 will respond with:

•  The same attention character “

:

” in ASCII or 

3A

 Hex

•  The address of the CLX-Ex2
•  The Reading
•  The Unit (mg/L)

A sample communication would look like this:
(Master computer requesting a report  

from address #1)  

:1 CRLF

(CLX-Ex2 set to address #1 Response)

  

:001 0.0249mg/L

9.1.2 Modbus Communication

Modbus protocol communication is included with this instrument.  

The Modbus address list is available from HF scientific or online at 

www.hfscientiifc.com.

Summary of Contents for HF scientific CLX-Ex2 Series

Page 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Page 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Page 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Page 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Page 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Page 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Page 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Page 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Page 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Page 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Page 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Page 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Page 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Page 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Page 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Page 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Page 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Page 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Page 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Page 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Page 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Page 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Reviews: