background image

8

4.2 Plumbing

The instrument is designed to require very little head pressure to operate, 

but will need around 0.34 bar (5 PSI). The maximum pressure for proper 

operation should not exceed 10.3 bar (150 PSI). The maximum 

allowable fluid temperature is 55°C (131°F).

NOTICE

The fluid waste from drain connection of this instrument contains 

reagents diluted with large quantities of sample water. HF scientific 

recommends that in the case of ballast installations, the waste 

stream is reintroduced to the bilge tanks. This waste fluid must 

NEVER be reintroduced into the incoming water stream.

All plumbing connections are intended to be made through stainless 

steel connections.
The water sample inlet requires 6 mm OD stainless steel tubing. The 

fitting is a compression fitting so the installer need only place the 

supplied the nut and feral over tubing and tighten.
The water drain requires 12 mm OD stainless steel tubing. This fitting is 

also a compression fitting so the installer need only place the supplied 

the nut and feral over tubing and tighten.
Properly installed and sealed connections are required to ensure the 

water tightness and ratings of the instrument.

NOTICE

The drain must be open to the Atmosphere for proper operation.

4.3 Air Supply

This instrument requires water and oil free air, (-40°C [-40°F ] Dew 

Point, Particles <5u, ISA Grade Hydrocarbon free). The air needs to be 

provided at between 5.5 and 7 bar (80-101.5 PSI) with a volume of 4 

cubic foot/min as measured at 5.5 bar (80 PSI). This air supply must be 

provided continuously from a known safe area. Min flow is 35 SLPM
The air connection requires 6 mm stainless steel tubing. A compression 

type fitting is supplied on the base of the main enclosure that simply 

requires the installer to slip the supplied nut and feral over the tubing 

and then tighten the nut.

4.4 Electrical Connections

All electrical connections must pass through the three compression 

grommets on the left side of the instrument. Cabling must be selected 

to ensure that these connections are tight. The grommets can accept 

cabling from 11 to 21 mm diameter. One grommet is supplied for 

system power and two are supplied for low voltage I/O connections. 

These grommets must be tightened to maintain cabinet air pressure.
Most of the low voltage electrical communication connections to the 

instrument are made at the termination area which is located on the 

left side of the instrument. Remove the access cover by loosening the 

captive screw. The connections are labeled and are self-descriptive. 

Please follow all government recommendations for installation of 

electrical connections to and between the instrument and other 

peripheral devices.
All terminals are designed to accept wires in the range of 14-28 AWG. 

All wires should be stripped to a length of 6 mm (¼").

For line power connections refer to the companion 

manual Catalog No. 28749.

4.4.1 RS-485

The RS-485 half-duplex (2-wire) digital interface operates with differential 

levels that are not susceptible to electrical interferences. This is why 

cable lengths up to 3000 ft can be implemented. The last device on 

each bus may require terminating with a 120 ohm resistor 

to eliminate signal reflection on the line. Do not run RS-485 

cables in the same conduit as power. Set-up of the RS-485 

is covered in 8.3 

Configuring the RS– 485 Port.

To prevent damage to the instrument, 

ensure that power is disconnected 

prior to making connections. For ease 

of connecting, remove the plug-in 

terminal block. Connections are 

labeled beside this termination on the 

PC board. The recommended cable 

is 22 AWG shielded twisted pair. The 

grey terminal block is removable to 

assist in making connections.

4.4.2 Relay

One relay alarm connections labeled 

ALM COM & ALM NO is provided. 

Please note that this is a fail- safe so 

the “normal” condition is power to the 

CLX-Ex2 on and no alarm condition. 

The relay is rated 240VAC 

2A. Operation of this alarm 

is covered in section 8.4

 

Configuring the Alarm.

4.4.3 4-20 mA

The 4-20 mA output is driven by a 15 VDC power source and can drive 

recorder loads up to 600 ohms. Transformer isolation is provided on 

the CLX-Ex2.

Do not run 4-20 mA cables in the same conduit as power. 

Operation of this output is covered in section 8.1 

Setting 

the 4-20 mA.

To prevent damage to the instrument, and for general safety 

ensure that power is disconnected to the CLX-Ex2 prior to making 

any connections. Polarities of the connections are labeled beside this 

termination on the PC board.
The recommended cable is 22 AWG shielded twisted pair. To prevent 

ground loops, connect the shield at either the CLX-Ex2 or at its 

destination, but not both. The grey terminal block is removable to assist 

in making connections.

The 4-20mA is factory calibrated. An adjustment is available 

on the 4-20mA in sections 8.12 and 8.13. In addition to 

making adjustments, these menus output continuous 4 mA 

or 20 mA and can be used as a signal test. Remember that 

the configuration mode will timeout after 15 minutes.

4.4.4 Remote Standby Connections

The instrument operates on user control software, where an external 

control initiates each cycle. This control line is called the RSI short for 

Remote Standby Input.
The CLX-Ex2 will indicate the current status on another line called RSO 

or Remote Standby Output. These connections are a relay contact that 

is closed when busy and open during when in standby mode.
The connections are labeled RSI and RSO. The RSI has + and – polarity 

connection associate with it and expects a 24VDC control signal.
The RSO is simply a relay contact and is not polarized.

Summary of Contents for HF scientific CLX-Ex2 Series

Page 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Page 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Page 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Page 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Page 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Page 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Page 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Page 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Page 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Page 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Page 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Page 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Page 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Page 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Page 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Page 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Page 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Page 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Page 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Page 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Page 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Page 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Page 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Page 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Reviews: