VTI Instruments Corp.
62
EX1629 Basic Operation
R
ESET
B
UTTON
The reset button on the rear panel of the EX1629, implemented according to the LXI LAN
Configuration Initialize (LCI) Mechanism specification, can be used to restore default network
settings. This is useful for recovery from an incorrect or unknown network configuration. To
perform a network reset:
1)
Power off the EX1629.
2)
Press and hold the reset button.
3)
Power on the EX1629.
4)
Continue to hold the reset button for at least 30 seconds.
5)
Release the reset button.
The EX1629 will power up as usual, but will use the default network configuration (DHCP and
AutoIP enabled) instead of its previous settings. VXI-11 Discovery (supported by the VISA
IO-Libraries) or the vtex1629_findinstr function can be used to determine the IP address of the
instrument.