VTI Instruments Corp.
220
EX1629 Command Set
vtex1629_self_cal_clear_stored
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_self_cal_clear_stored (ViSession
vi
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function clears self-calibration data from nonvolatile memory. It does not, however, affect the current self-
calibration data, regardless of whether it came from a recent self-calibration or a loading from nonvolatile memory
upon power-up initialization. Thus, the effective calibration will not be changed by this call. To return to an
operating state where only the full calibration is used, the vtex1629_self_cal_clear function can be called, or the
system can be rebooted.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
…
status = vtex1629_self_cal_clear_stored (instrumentHandle);