VTI Instruments Corp.
66
EX1629 Triggering
S
YNCHRONIZING
M
ULTIPLE
I
NSTRUMENTS
In cases where larger acquisition systems are required, multiple devices can be configured to
utilize the same ADC sample clock and synchronization signal to acquire data that is synchronized
across all acquisition channels. This requires configuring one of the devices as a master and the
remaining devices as slaves. The devices must be connected together using the LXI Trigger Bus in
either a star or daisy chain configuration. The details of how to program the configuration of
multi-box systems is described in Section 6.
Configure Trigger and ADC Clock
for further details.