www.vtiinstruments.com
EX1629 Command Set
219
vtex1629_self_cal_clear
F
UNCTION
P
ROTOTYPE
ViStatus vtex1629_self_cal_clear (ViSession
vi
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
D
ATA
I
TEM
R
ESET
V
ALUE
Not applicable to this function.
D
ESCRIPTION
This function clears the current self-calibration data. This operation clears the volatile data, but does not affect any
self-calibration data that is stored in nonvolatile memory. After this function, the effective calibration will be the full
calibration (factory calibration, also known as the annual calibration) plus, if it exists, the nonvolatile self-
calibration. If a self-calibration is performed, and non-volatile self-calibration data exists, executing this function
effectively reverts to the non-volatile self-calibration data. With no non-volatile self-calibration data, this effectively
reverts to full (factory/annual) calibration data.
E
XAMPLE
ViSession instrumentHandle;
ViStatus status;
…
status = vtex1629_self_cal_clear (instrumentHandle);