98-100 ISSUE 8
14
U
SING THE
DC1000
AND A
V
OLTECH
AT3600/AT5600
WITH A SAFETY SYSTEM
When using your DC1000(s) with a Voltech AT a bespoke “Y” lead should be used
for the interlock system
Fig. 25 shows this connection.
Fig. 25
INTERLOCK
A) Use the Safety Device “Y” lead (250-031).
The end labeled
“SD
” connects to the Safety Light Curtain Interface cable (250-030)
from the Control Box (
1)
.
The end labeled
“AT”
is connected to the AT Safety Interlock (
2).
The end labeled
“DC”
is connected to Safety Interlock IN (
3)
of DC1000A #1.
B. If using more than one DC1000, then use a 9way-9way M-F straight through lead
(77-046) to connect Safety Interlock OUT (
4)
of DC1000A #1 to the Safety Interlock
IN (
5)
of DC1000A #2. This can be repeated for up to 20 DC1000s in a chain.
COMMUNICATIONS
C. Use a 9way-9way M-F straight-through lead (77-046) to connect AT Aux (Port
6)
to RS232 In (
7)
of DC1000A #1.
D. Use a 9way-9way M-F straight-through lead (77-046) to connect RS232 OUT (
8)
of DC1000A #1 to RS232 IN (
9)
DC1000A #2. This can be repeated for up to 20
DC1000s in a chain.
E. Use 9way-9way F-F RS232 lead (77-015 AT Editor Cable) to connect RS232 OUT
(
10)
of DC1000A #2 to PC RS232 (
11)
.