98-100 ISSUE 8
32
“D_COUNT,xx <cr> <lf>”
where
xx
is the number of units in the chain.
Within 2 seconds, the unit will respond with:
“D_STAT,0,
result
<cr> <lf>”” where
result
is a single integer indicating the
status of the unit and any other unit attached.
If everything is OK, 0 or 1 will be returned. If not, then a status number will
be returned.
S
ET
DC1000 O
UTPUT
C
URRENT
Command
“D_SET,
level
<lf>”, where
level
is the demand in milli-Amps and can range
from 100 to 25000. If multiple units are connected together, then the same
current will be sent to each unit.
Messages returned
The following messages will be returned:
Within 100ms, an attached unit will respond with:
“D_COUNT,xx <cr> <lf>”
where
xx
is the number of units in the chain.
Within 2 seconds, the unit will respond with:
“D_STAT,0,
status
<cr> <lf>” where
status
is a single integer indicating the
status of the unit and any other unit attached.
If everything is OK, 0 or 1 will be returned. If not, then a status number will
be returned.
G
ET THE STATUS OF THE UNIT
Command
“D_STAT? <lf>”
This command will return the status of the attached DC Bias unit