9.Technical Specifications
113
9.
Technical Specifications
Unless otherwise specified, the technical specifications applied are for DSO
four-channel series only, and Probes attenuation set as 10X. Only if the oscilloscope
fulfills the following two conditions at first, these specification standards can be
reached.
This instrument should run for at least 30 minutes continuously under the
specified operating temperature.
If change of the operating temperature is up to or exceeds 5
℃
, do a
"Self-calibration" procedure (see "
How to Implement Self-calibration
All specification standards can be fulfilled, except one(s) marked with the word
"Typical".
Oscilloscope
Performance Characteristics
Instruction
Bandwidth
DSO-6084 80 MHz
DSO-6104 100 MHz
DSO-6204 200 MHz
Vertical Resolution (A/D)
8 bits
Channel
4
Waveform Capture Rate
DSO-6084
DSO-6104
45,000 wfms/s
DSO-6204 70,000 wfms/s
Multi-level Gray Scale Display &
Color Temperature Display
(Use gray scale to indicate frequency of
occurrence, where frequently occurring
waveform are bright.)
Support
Acquisition
Mode
Normal, Peak detect, Averaging
Sample rate
(real time)
DSO-6084
DSO-6104
Four CH 250 MSa/s
Dual CH* 500 MSa/s
Single CH 1 GSa/s
DSO-6204
Four CH 500 MSa/s
Dual CH* 1 GSa/s
Single CH 1 GSa/s
Input
Input coupling
DC, AC , Ground
Input impedance
1 MΩ±2%, in parallel with 15 pF±5 pF
Input coupling
0.001X - 1000X, step by 1 – 2 - 5
Max input voltage
400 V (DC + AC Peak)
Bandwidth limit
20 MHz, full bandwidth