18
EMC GUIDANCE
COMPILED EUROPEAN STANDARDS LIST
Risk Management
EN/ISO 14971:2007
Labeling
EN 980:2008
User Manual
EN 1041:2008
General Requirements
for Safety
EN 60601-1:2006/AC:2010
EN 62304:2006/AC:2008
EN 60601-1-6:2010
EN 60601-1-11:2010
Non-invasive
Sphygmomanometers
General Requirements
EN 1060-1:1995+A2:2009
EN 1060-3:1997+A2:2009
EN 1060-4:2004
Electromagnetic
Compatibility
EN 60601-1-2:2007/AC:2010
EMC GUIDANCE
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions for all EQUIPMENT
and SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic emission
The VGP-4300 is intended for use in the electromagnetic environment specified below. The
user of the monitor VS-4300 should assure that it is used in such an environment.
RF Emissions
CISPR 11
Group 1
The VGP-4300 must emit electromagnetic
energy in order to perform its intended
function. Nearby electronic equipment may
be affected
RF Emissions
CISPR 11
Class B
Harmonic Emissions
IEC 61000-3-2
Not
Applicable
Voltage Fluctuations / Flicker
Emissions IEC 61000-3-3
Not
Applicable