20
EMI GUIDANCE
Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The TMB-1014 is intended for use in the electromagnetic environment specified below.
The customer or user of TMB-1014 should assure that it’s used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
N/A
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the TMB-1014, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.167
d = 1.167 80 MHz to 800 MHz
d = 2.333 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: