EMC GUIDANCE
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity – for all ME
EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The VGP-4300 is intended for use in the electromagnetic environment specified below. The
user of the VGP-4300 should assure that it is used in such an environment.
Immunity Test
IEC 6060 1
Test Level
Compliance
Level
Electromagnetic environment
- guidance
Electrostatic discharge
(ES D) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic time. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%
Electrical fast transient
/ burst IEC 61000-4-4
±2 kV for
power supply
lines
N/A
Surge IEC
61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s) to
earth
N/A
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% UT
(>95% dip in
the UT)
for 0.5 cycles
N/A
40% UT
(60% dip in
the UT)
for 5 cycles
N/A
70% UT
(30% dip in
the UT)
for 25 cycles
N/A
<5% UT
(>95% dip in
the UT)
for 5 cycles
N/A
Power Frequency
(50Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a commercial or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test level