
43
C. OPERATION
C.1 PURGING THE SYSTEM
C.1.a Air Bubbles
The presence of air bubbles in any of the detectors will greatly increase the level of
baseline noise in the signal. It is imperative that bubbles be kept out of the detectors. There are
basically two ways for air bubbles to get into the system: Purging is accomplished by operation
of the Refractometer and Viscometer Purge valves as previously described.
1.
By Entry from the Outside
This ordinarily happens only when the detector is disconnected and then reconnected.
Air may enter by siphoning of solvent from the detector, or from air in the connecting tubing.
To minimize this, always immediately plug off the Sample In port when disconnecting the
connecting tubing. Then before reconnecting the tubing, be sure to purge the tubing of any
air that may have entered.
2.
By Spontaneous Bubble Formation in the Solvent
If the solvent is not thoroughly degassed before use, it can occasionally form bubbles
spontaneously on the wall of any static cavity. The best way to prevent this from happening
is to use only filtered / vacuum degassed mobile phase solvent. For this purpose,
an in-line
solvent degasser before the HPLC pump is most convenient and highly
recommended.
Air bubble problems seldom occur in systems operating with an in-line
degasser. Sparging is not a recommended method for degassing solvent.
C.1.b Changing Mobile Phase Solvent
The other reason for purging is upon changing mobile phase solvent composition.
1. Be certain that the new mobile phase is miscible with the old solvent that is in the detector.
In most cases, the detector is shipped in THF, which is miscible with virtually all organic
solvents and also with water. However, it is not miscible with all aqueous salt solutions.
The best practice when going from THF to an aqueous salt solution is to go through an
intermediate flush with distilled/deionized water.
2. To make a change of mobile phase in the columns, disconnect the detector from the system
while making the solvent changeover in the columns. Reconnect the detector only after the
columns are sufficiently purged and equilibrated with the new mobile phase. The reason for
this is that columns frequently discharge residues upon solvent changeover and these
residues can sometimes coat the capillary walls of the viscometer bridge and deposit debris
on the surface of light walls causing high background and noise.
3. To replace the columns with another set that is already equilibrated to the new mobile phase
solvent: