Components
CAPILLARY BRIDGE
The four capillaries R
1
- R
4
are stainless steel 1/16”od x 0.01”id x 24”l. They are
wrapped together in a harness and suspended in the center panel module. The inlet tubing is
also 0.01”id tubing. Most other tubing connections to the bridge (to transducers, purge ports,
viscometer exit tubing) are 0.04”id,
Figure 15
.
Figure 15: Viscometer components
PRESSURE TRANSDUCERS
The DP and IP transducers are of the magnetic reluctance type. They are mounted on
the inside center panel of the detector module. Connections via 0.04”id ss tubing go to the
transducers and to the purge valves. The IP transducer is positioned in series with the RI
reference purge and is thus purged with the RI.
PURGE SOLENOIDS
The transducers must be purged completely of air bubbles in order to obtain an optimum
baseline. This is accomplished by using the purge switches located on the front panel while
flow is on and allowing solvent to drain to waste for a few minutes.
DP Solenoids
RI/IP Solenoid
Bridge Capillaries
IP Transducer
DP Transducer
Delay Columns
Filter
Pulse Dampener