B. INSTALLATION
B.1 INSTALLATION KIT FOR MODEL 302 TDA
Quantity
Item
1 Power
Cable
1 12-Pin
Phoenix
Connector
1 Viscotek
Screwdriver
1
Allen Head Wrench, 0.050"
1
36" SS Tubing, 1/16"x 0.01"id
1
36" TFE Tubing, 1/16"x 0.01"id
1
36" TFE Tubing, 1/16"x 0.03"id
10
Valco
Nuts,
1/16"
10
Valco
Ferrules,
1/16"
3 Valco
Cap,
1/16"
2 Valco
Plug,
1/16"
1
Valco Union, 1/16”x 0.03”id
1 RALLS
Filter:
2
Teflon
O-ring P/N
FIL060
2 SS Frit - 0.5 micron
P/N 728025
10
Filter
Membranes:
5
-
2046
nylon
(Aqueous)
5
-
2013
PTFE
(Organic)
1
PS 90 Narrow Calibration Standard
1
PS 280 Broad Calibration Standard
1 Instrument
Manual
2
4 Ft. Analog Signal Cable (1 Pr..Shielded)
1
10 Ft. Trigger Cable (1 Pr.)
1
10 Ft. Serial Cable
1 10 Ft. Diagnostic Cable
B.2 CONNECTION TO SEC SYSTEM
The Model 302 TDA is designed to be directly to the outlet of a properly functioning
HPLC pump by the minimum length possible of 0.01”id SS tubing.
1. Check the power receptacle on the back and make sure that the voltage set is correct for
the AC power source (110V/220V) being used. Connect the power cord and turn on the
main power switch on.
2. The detector is normally checked out and shipped with THF solvent. If possible, it is best to
use THF for the initial checkout at the site. If another mobile phase is to be used, be certain
that it is miscible with THF before proceeding.
3. With the detector drawer removed from the oven module, flush fresh solvent from the pump
to the connection provided to the inlet of the separation columns.