52
MPOLx-85 / MPOLx-85P
7 PCM & PROBE O
PERATION
O
PERATION
Overlay
J
Press the
[
A
]
key and then the
[
Overlay
]
button to change the
overlay view. Repeat the action until the desired view appears:
without overlay > with colored edges > with colored edges
and legend > without overlay > ...
A summary of test results is displayed at the upper right of the
screen (following figure is just an example and may vary).
If the test has been passed, the image has a green colored frame.
If failed, the frame is red.
Fig. 18 Pass/Fail analysis – overlay view
(figure is showing a simplex fiber)
To return to the live image:
J
Press the QuickCapture™ key.
– or –
Press the
[
A
]
key and tap the
[
Live
]
button.
Saving test results
The image can easily be saved.
Zone A:
Represents the core zone. It is the area surrounding
the core.
Zone B:
Represents the cladding zone. It surrounds the
majority of the fiber cladding.
Zone C:
Represents the epoxy ring.
Zone D:
Represents the ferrule zone. It identifies a portion of
the ferrule near and around the fiber.
Zone D
Zone C
Zone B
Zone A