5.3 Cleaning and Disinfecting the probe
Cleaning the probe
The operation process is as follows:
The following cleaning methods shall be taken immediately after each complete test on the
patient to avoid drying of the coupling remaining on the probe, which is not conducive to the
cleaning
of the probe and to prevent the body and sound head from being corroded by coupling or other
reagents . Before the cleaning or disinfection work begins, the operator should wear
professional
protective equipment such as goggles and gloves, and check whether the instruments,
equipment,
materials and environment used meet the requirements in advance.
1
、
Turn off the power of the ultrasound probe, press and hold the button for more than 3s until
the
light goes out. For the specific operation process, please refer to the product instruction
manual.
At the same time, check whether the waterproof cap of the charging port of the machine is
installed and returned. If it is opens, the waterproof cap should be installed and returned.
2
、
Use clean special wiping paper for ultrasonic instrument (the wiping paper shall be soft non
abrasive disposable paper towel) or clean lint free soft cloth to wipe the acoustic head. Gently
wipe the paper or soft lint-free cloth along the wide side of the vocal head to the other end ( if
the
wiping force is too large, it may cause the depression of the vocal head and make the machine
unusable) . The wiping direction should be in accordance with the Repeated wiping is
prohibited
in the same direction. If the sound head is also stained with coupling or other reagents, it
should
also be wiped in the same way.
3
、
During the wiping process, after each wiping from the beginning to the end, fold the special
wipe
paper for ultrasonic instruments in half, and fold the wipe paper in half towards the side stained
with coupling or other ultrasonic special colloid residues (Fold the stained side inside the
wiping
paper).
4
、
Ultrasound instrument special wipes or lint-free soft cloth should not be folded in half at most
for
more than 4 times. If there is still coupling or other ultrasonic special colloids remaining on the
ultrasonic probe part after wiping for the 4th time, you need to replace with new ultrasonic
instrument special wipes, and repeat after replacement. The above steps are performed until
the
coupling remaining in the vocal head is completely wiped.