1.
Matters needing attention in clinical examination technology
:
This equipment can only be operated by qualified
medical personnel.
This manual does not introduce a clinical examination
technique. It is necessary to select the correct inspection
techniques according to the professional training
knowledge and clinical experience.
2.
The equipment can not be checked for a long time.
3.
Do not use incompatible coupling agents, disinfectants, probe
protective cover, probe, puncture rack.
4.
Sterile gloves must be worn to prevent infection when using
ultrasonic probes.
5.
You must use a sterile ultrasound coupling agent. Use a coupling
agent that is in compliance with local regulatory requirements. In
addition, it is necessary to properly manage and use the
ultrasonic coupling agent to ensure that it does not become a
source of infection.
6.
The probe cover is made of natural rubber and is used with
caution for natural rubber allergy.
7.
For in vivo transducers in a single fault condition, the surface
temperature rise shall not exceed 43 ° C.
CAU
TIO
N
:
1
In order to prevent abnormal probe function, read the following safety
precautions:
After each ultrasonic examination, the ultrasonic coupling agent on the
surface of the probe should be thoroughly erased. Otherwise, the
ultrasonic coupling agent will be solidified on the probe head, which
will affect the quality of the ultrasound image.
The probe should be cleaned and disinfected before and after each
ultrasonic examination
。
2
Ambient environmental requirements:
Please use the ultrasonic probe in the specified environment
:
ambient temperature
:
0
℃
~
35
℃
relative humidity
:
30%
~
85%
(
No condensation
)
Atmospheric pressure
:
70KPa
~
106KPa
。
To prevent damage to the ultrasonic probe, do not expose the probe to
the following environment
:
Place where the sun shines
A place where the temperature changes dramatically.
A place filled with dust
Easy to vibrate place
Place near the heat source
3
Repeated disinfection will lead to the safety and performance of the
probe, the performance of the probe should be regularly checked.