CE-C10-03-SMSP01-20220502 V1.0
39
5
、
Use ultrasonic special wiping paper or lint free cloth dipped with enzyme containing
detergent(the usage of specific cleaners should refer to the usage method of cleaner labels
and
instructions, and pay attention to whether the use time of detergent is within the warranty
period
of the product) to clean until there is no visible stain on the acoustic head and the acoustic
head and its surroundings have been wiped.
6
、
Rinse the probe with drinking water or softened filtered water ( the water quality of tap water
is
too hard, there may be residual salt and alkali substances, which will cause different degrees
of
corrosion to the ultrasonic probe. For specific water quality requirements, please refer to AAMI
TIR 34 ) . During the process, keep wiping the vocal head with your hands until all the cleaning
agents are rinsed. In principle, the rinsing time should not be less than 30s. During the rinsing
process, water should be prevented from splashing on the fuselage. If any liquid enters the
fuselage,it will cause irreversible damage to the electronic components of the machine.7
、
Use
the special wipe paper for ultrasonic instruments or a clean lint-free soft cloth (The same lint
free soft cloth used before is no longer used) to wipe the entire body and the vocal head, and
wipe the residual water stains on the acoustic head until all the residual water stains on the
body
are wiped clean.
8
、
Place the whole ma
chine in the air with humidity ≤ 60 % , temperature ≤ 40
℃
and ≥ 15
℃
,
fine
particles ≤ 3 5ug/ m³ Ventilation and drying shall be carried out in a ventilated indoor
environment,, and the drying time should be ≥ 3 0 s. Meanwhile, direct exposure to sunlight
shall
be avoided.
9
、
Check whether the whole part of the fuselage is damaged, such as cracks, cracks, cracks or
protruding sound head.In case of the above situations, stop using immediately and contact the
local dealer or sales representative.
Disinfecting the probe
The operation process is as follows:
Before the machine is sterilized, it should be ensured that the cleaning work has been carried
out and the cleaning process is correctly followed. Operators should wear protective
equipment such
as professional goggles and gloves, and check whether the instruments, equipment, materials
and
environment used meet the requirements in advance.
1
、
Use alcohol disinfectants (For example: The compound double- chain quaternary
ammonium
salt with free
safety content ≤ 2% is mixed with purified water or softened filtered water. The