Manual for:
VGE Pro UV Control 400/420/600 and VGE Pro UV Control 140/200
EP600001, EP200002
’main menu’
à
’settings’
à
’UV sensor’
contains all UV sensor specific settings parameter
If no sensor is activated only
the menu item ‘number of sensors’ will appear in ‘UV sensor’ sub menu.
exit
- Exit menu
100% value
(default 100W/m²)
set the 100% reference value for each sensor
• this value is used for displaying UV and alarm/warning thresholds in %
• defines the reference when forwarding the UV values using the 4..20mA
output. Note:
• if more than one sensor is present an additional submenu allows sensor selection;
• the 4..20mA output represents the lowest relative UV value of all sensors connected
warn value
(default 80W/m²)
UV warning threshold for activating warning relay on base unit and yellow LED 'warning' on frontend
alarm value
(default 60W/m²)
UV alarm threshold for activating alarm relay on base unit and red LED 'alarm' on frontend
calibration factor
(default 100%)
adjust the displayed UV value (on deviations from a reference sensor)
address
…
enter submenu for setting bus address
’settings’
à
’UV sensor’
à
‘address..’
exit
- Exit menu
scan bus
scans the sensor bus. The sensor address will be detected and entered automatically in sub
menu
‘address’ (use only for one digital sensor connected)
address
enter the address for digital UV sensors manually (if one or more digital sensor is connected)
number of sensors
(default 0)
set number of connected sensors;
set to “0” to deactivate UV monitoring via digital sensors
warn delay
(default 10 sec.)
set appropriate delay