V1410 / V1412 User Manual
98
Problems in EtCO2 Measurement (Microstream)
Problem
Remedy
FILTER LINE NOT
CONNECTED Message
displays.
The filter line is not connected to the main body.
Connect filter line.
EtCO2 waveform and data
are not displayed.
1. Verify the sample line is not occluded.
2. Change the sample line if there is questionable
operation or possible leaks in the line.
Problems in EtCO2 Measurement (Mainstream/Sidestream)
Problem
Remedy
SENSOR DISCONNECT
message displays.
The filter line is not connected to the main body.
Connect filter line.
EtCO2 waveform and
numeric are not displayed.
1. Verify the sample line is not occluded.
2. Change the sample line if questionable operation
or possible leaks in the line.
Problems in the Printer (Optional)
Problem
Remedy
No printing.
1. Check if the message PRINTER is displayed on the
screen.
2. Check if the power LED lights up on the printer.
3. Check if the paper is properly loaded.
4. Check if the printer door is closed.
Noise during printing.
1. Check if the roller is contaminated with foreign
materials or if the gear is out of position.
2. Check if the problem is due to a paper jam. Change
the paper if necessary.