V1410 / V1412 User Manual
56
2
Attach a patient’s electrode to the sites shown below.
<Standard 3 Electrode Placement>
IEC
AAMI
Attaching site
Lead
Electrode
Color
Lead
Electrode
Color
R
Red
RA
White
Just below the clavicle (scapula) near the right shoulder
L
Yellow
LA
Black
Just below the clavicle (scapula) near the left shoulder
F
Green
LL
Red
Lower area of the left side of the abdomen
N
Black
RL
Green
Lower area of the right side of the abdomen
C
White
V
Brown
Select the candidate lead for measurement among the chest
leads
NOTE
The pulse may be chosen from the ECG or SpO2 sensor if both are on. If one of the
sensors is off, then the pulse reading from the remaining sensor is displayed in both ECG
and SpO2 areas of the display.
The default setting of ECG Filter is ANIMAL.
The default setting of ECG Lead is Lead II.
When using an ECG 5-lead cable, you need to set the Menu for a 5 lead ECG cable.
Caution
Be certain the correct setting within the ECG configure-cable menu is chosen to match the
cable you are using (3 or 5 lead). If the wrong cable is chosen in the menu settings, the
monitor will be unable to display any ECG information.