![Vetland V1410 User Manual Download Page 71](http://html.mh-extra.com/html/vetland/v1410/v1410_user-manual_1010098071.webp)
V1410 / V1412 User Manual
71
Changing the NIBP Unit of Blood Pressure
1
To change the unit of blood pressure, select UNIT
in the NIBP Menu and set the desired unit.
Setting Mode
Range
Description
MODE
AUTO, MANUAL,
VENOUS STAT, and STAT
Set the automatic measurement.
INITERVAL
2, 3, 4, 5, 10, 15, 30 min,
1, 2, 4, 8, 12 Hours
Set the interval of the automatic NIBP
measurement time.
STAT
5 –15 min (1 min interval)
Set the STAT MODE measurement time.
SCREEN: Set ON/OFF
PRESSURE
MAX LIMIT, VENOUS STAT
MAX LIMIT: Limit value for pressure
VENOUS STAT: Pressure value of
VENOUS STAT MODE
LEAKAGE
Engineer Mode
UNIT
mmHg, kPa
Set the unit of blood pressure.
Caution
Max Limit must be matched to the patient and set above anticipated systolic pressure to allow
for proper reading.
Note
MAX LIMIT can be set between:
–
VET_Large: 100 – 300 mmHg (5 mmHg unit)
–
VET_Small: 70 – 300 mmHg (5 mmHg unit)
Default setting of the MAX LIMIT is 280 mmHg.