Electrical Safety Precautions
Please read the following before using the system:
CAUTION
To prevent electric noise during use, the product should be installed apart
from any dynamo, X-ray equipment, broadcast equipment, or portable
cables. An inaccurate result may occur if this equipment is placed near the
product.
When not in use for extended periods of time the battery should be
disconnected to prevent excessive depletion.
Damaged cords or cables may cause fire or electric shock.
NOTE
The V1410 / V1412 is:
-
Class IIb equipment.
-
Type CF grade for ECG, Resp, and IBP.
-
Type BF grade for EtCO2, SpO2, NIBP, and Temperature
The noise level is Class B according to the IEC/EN 60601-1 (Safety of
Electric Medical Products).
The noise redemption is Level B according to the IEC/EN 60601-1-2
(Electromagnetic Compatibility Requirements).
The protection grade against water is IPX-0.
Patient Safety Precautions
To protect the patient observe the following precautions:
CAUTION
The analysis of arterial blood pressure measurements should be
judged by qualified veterinary professionals.
Arterial blood pressure may be influenced by a patient’s posture,
physical condition and other factors such as a patient’s movement.
Premature patients and patients with chronic pulmonary disease
should be checked for oxygenation levels before starting treatment.
Sensor-attachment sites should be checked at least every 4 hours.
Please check if the sensors are properly attached, if the skin condition
is normal, and if the sensors are well positioned. Special care is
required for patients with poor blood flow. If the sensor attachment
status is not checked periodically, it may lead to skin injury by
extended contact and necrosis from the pressure. Patients with weak
blood flow should be checked regularly so that injury cannot occur.