![Vetland V1410 User Manual Download Page 48](http://html.mh-extra.com/html/vetland/v1410/v1410_user-manual_1010098048.webp)
V1410 / V1412 User Manual
48
Parameter
Item Name
VET_Large
VET_Small
TEMP
TRACE
CH 1
TRACE ON
TRACE ON
CH 2
TRACE ON
TRACE ON
CH 3
TRACE OFF
TRACE OFF
CH 4
TRACE OFF
TRACE OFF
UNIT
◦
C
◦
C
ALARM
HIGH
39.0
◦
C
39
◦
C
LOW
36.0
◦
C
36
◦
C
TYPE
HIGHEST
HIGHEST
RESP
SIZE
X1
X1
SPEED
6.25 mm/sec
6.25 mm/sec
TRACE
TRACE ON
TRACE ON
ALARM
HIGH
60 bpm
60 bpm
LOW
5 bpm
5 bpm
TYPE
HIGHEST
HIGHEST
RESP APNEA TIME
OFF
OFF
NIBP
MODE
MANUAL
MANUAL
INTERVAL
30 MIN
30 MIN
STAT
TIME
5 MIN
5 MIN
SCREEN
OFF
OFF
PRESSURE
MAX LIMIT
280
280
VENOUS STAT
160
100
UNIT
mmHg
mmHg
TREND
SCREEN (OFF)
SCREEN (OFF)
ALARM
HIGH
160/90 (110)
160/90 (110)
LOW
90/50 (60)
90/50 (60)
TYPE
HIGHEST
HIGHEST
EtCO2
Gain
x2
x2
Speed
6.25 mm/sec
6.25 mm/sec
UNIT
mmHg
mmHg
TRACE
TRACE OFF
TRACE OFF
ALARM
HIGH
40
40
LOW
10
10
TYPE
HIGHEST
HIGHEST