EtCO2 (End-Tidal CO2) – Optional
•
Microstream
- Measurement Mode: Microstream
- Range: 0 – 99mmHg
- Temperature: 0 – 50
℃
- Accuracy:
Time
* CO
2
Density
Accuracy
0 – 20 min
0–38 mmHg
±4 mmHg
39–99 mmHg
±12%
> 20 min
0–38 mmHg
±2 mmHg
39–99 mmHg
±5% of r 0.08% for every
1 mmHg (above 38 mmHg)
* (760 mmHg) level
- Respiration Measurement Range: 0 – 150 bpm
- Respiration Accuracy: 0 – 70 ±1 bpm
71 – 120 ±2 bpm
121 – 150 ±3 bpm
- Initial Time: Self test for 30 seconds after turning on the device.
- Sampling Value: 50 ±7.5 ml/min
- Period of Calibration: After 1 year or 4,000 hours of operation
- Response Time: 2.9 seconds after measurement (including rising time and delay time)
Rising Time – 190 msec, Delay Time – 2.7 sec
•
Mainstream
- Initial Time: less than 15 seconds (25
◦
C)
- CO
2
Measurement Range: 0 – 150 mmHg, 0 – 19.7%,
0 – 20 kPa (at 760 mmHg)
- CO
2
Resolution: 0 – 69 mmHg
0.1 mmHg
70 – 150 mmHg
0.25 mmHg