39
■
Removal
Removal
Be sure to secure environmental and personnel safety before removing the ion bar.
CAUTIONS
UNPLUG
Be sure to unplug the power supply cable when removing the ion bar to
avoid an electric shock or any accident caused by abnormal operation.
REMEMBER
Ensure that the air supply is stopped before disconnecting the air hose.
Keep your face apart from the air hose connector when disconnecting the
air hose because compressed air may spout out from the air hose.
Disposal
Disposal
Follow the law and regional regulations when abolishing the ion bar.
Troubleshooting
Counteractions against abnormalities
Be sure to unplug the power supply
cable before troubleshooting an abnormality.
Contact the distributor of the ion bar when repair is necessary.
Abnormality
Cause and inspection
Counteraction
Static electricity
cannot be
removed.
Check if the electrode needle is dirty.
Clean the electrode needle.
Check if there is any alarm.
Take an action according to the alarm.
Check if there is any conductive material
around the ion bar.
Keep any conductive material
apart from the ion bar.
Check if the power supply cable length is
10 m or shorter.
If the power supply cable is too
long, the ion bar cannot operate
normally due to voltage drop.
Performance of
static elimination
is lowered.
Check for the correct frequency setting.
Check for the correct duty ratio setting.
Check for the correct voltage level setting.
Check if there is any conductive material
around the ion bar, or if the adjacent erasers
are too close.
Alarm is output
frequently.
Check if the electrode needle is dirty.
Clean the electrode needle.
Check if there is any conductive material
around the ion bar.
Keep any conductive material
apart from the ion bar.
Signal output
failure
Check the cables.
Use Category5E cable(Category
6 or 7 can be substituted).
Remote
controller failure
Check the distance between the ion bar
and remote controller.
Use the remote controller within
8 m from the ion bar.
Check for the battery.