Operations & Maintenance Manual: Important Information
0900‑4868 REV‑08
QuickConnect Cable, Monitor, and Cart Only: This product may only be cleaned or disinfected
by using the approved processes provided in the GlideScope and GlideRite Products
Reprocessing Manual (part number 0900‑5032). Cleaning and disinfection methods listed are
recommended by Verathon based on efficacy or compatibility with component materials.
WARNING
QuickConnect Cable, Monitor, and Cart Only: For information on the handling and disposing of
recommended reprocessing solutions, please refer to the solution manufacturer’s instructions.
WARNING
Power Adapter Only: Electric shock hazard. Do not immerse the power adapter in water. When
cleaning the power adapter, use a cloth dampened with isopropyl alcohol on the outside of the
enclosure.
WARNING
WARNINGS: PRODUCT SAFETY
This instrument and related devices may contain batteries or other environmentally hazardous
materials. When the instrument or accessories have reached the end of their useful service life,
see the section Device Disposal. Dispose of used, single‑use components as infectious waste.
WARNING
Two areas of the bronchoscope tip that contact the patient can exceed 41°C (106°F) as part of
normal operation:
The first area is the light‑emitting area surrounding the camera in the tip. When used as
indicated, continuous contact with this area is unlikely because, if tissue were to contact this
area, a usable view would be lost. Devices would then need to be adjusted to regain the
working distance needed for a usable view.
The second area is the area of the tip that surrounds the camera, but is out of its field of view.
Continuous contact with this area is unlikely because the product is typically not held stationary
for an extended period of time, and because there is normally a small clearance between the
area and the adjacent tissue.
To prevent thermal damage, such as a burn to the mucosal tissue, avoid prolonged, continuous
contact with these areas of the bronchoscope tip.
WARNING