Operations & Maintenance Manual: Important Information
0900‑4868 REV‑08
CAUTIONS
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and must be installed and operated according to the instructions in this
manual. For more information, see the Electromagnetic Compatibility section.
Avoid using the GlideScope system adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, observe the system to verify normal operation in the configuration in
which it will be used.
This device can radiate radio frequency energy and is highly unlikely to cause harmful
interference with other devices in the vicinity. There is no guarantee that interference will
not occur in a particular installation. Evidence of interference may include degradation of
performance in this device or other devices when operated simultaneously. If this occurs, try to
correct the interference and restore optimal image quality by using the following measures:
• Turn devices on and off in the vicinity to determine the source of interference
• Reorient or relocate this device or other devices
• Increase the separation between devices
• Power off and on the monitor if the image quality is not optimal after removal of the
interference
• Connect the device to an outlet on a circuit different than the other device(s)
• Eliminate or reduce EMI with technical solutions (such as shielding)
• Purchase medical devices that comply with IEC 60601‑1‑2 EMC standards
Be aware that portable and mobile radio frequency communications equipment (cellular phones,
etc.) may affect medical electrical equipment; take appropriate precautions during operation.
CAUTION
Monitor Only: The system contains electronics that may be damaged by ultrasonic and
automated washing equipment. Do not use an ultrasonic device or automated washing
equipment to clean this product.
CAUTION
Risk of permanent equipment damage. This product is sensitive to heat, which causes damage
to the electronics. Do not expose the system to temperatures above 45°C (113°F), and do not
use autoclaves or pasteurizers. Use of such methods to clean or disinfect the system causes
permanent device damage and voids the warranty. For a list of approved cleaning procedures
and products, see the GlideScope Products Reprocessing Manual (part number 0900‑5032).
CAUTION