ELGA
PURELAB Option-S 7/15 US
Operator Manual
PURELAB Option-S 7/15 US Version 1 08/14
Page 19
6. OPERATION
The
PURELAB Option-S US
will run automatically and will signal
alarm conditions to ensure prompt efficient system management and
corrective action.
Note:
To ensure that water purity remains at a
high level it is important to leave the unit
in process mode.
6.1 Alarm Conditions
Alarms will signal at the following conditions:
Replace Ion-exchange Cartridge
The Ion-exchange cartridge replacement alarm is signalled by
an audible alarm and flashing icon at the default setting of
4380 hours (6 months) of use. Press the button to mute the
audible alarm. Follow the instructions to replace the Ion-
exchange cartridge.
(See Section 7.2)
.
Replace Pre-treatment Cartridge
The pre-treatment cartridge replacement alarm is signalled by
an audible alarm and flashing icon at the default setting of 4380
hours (6 months) of use. Press the button to mute the audible
alarm. Follow the instructions to replace the pre-treatment
cartridge.
(See Section 7.1)
.
Reservoir Level Disconnect Fault Alarm
The reservoir level disconnect fault alarm condition will signal
with an audible alarm and flashing icon. Press the button to
mute the alarm. This alarm condition will cause the process to
turn off. Connect the reservoir level and press the process key
to clear the alarm.
Water Purity Alarm
This alarm will signal if the water purity deviates from the preset
parameters and will cause the water purity value to flash and an
alarm to sound, until water purity improves to within acceptable
purity limits. Press the button to mute the alarm. If water
purity stays outside acceptable purity limits replace the ion-
exchange pack following the instructions in section 7.2.
Flash
Normal Process Screens
Flash
Water Purity Alarm / Standby
Alarm Conditions
Reservoir Level
Disconnect Fault