PURELAB Option-S 7/15 US Version 1 08/14
Page 13
Outlet to
reservoir
Blank
Blank
Drain
Outlet and Drain Connections
5.3 Connecting up the PURELAB Option-S US
Once the
PURELAB Option-S US
unit has been positioned either on
a wall or on a bench, it should be connected as follows:
•
Mains water inlet tube
•
Drain
•
Outlet to reservoir
Step 1 - Fitting Tubes
1. PUSH in collet on connector.
2. PULL out transit plug.
3. CUT a clean square end on an 8mm OD semi
rigid drain tube.
4. PUSH tube into connector.
CAUTION!
Do not restrict drain line.
CAUTION!
If the water supply is at a pressure greater
than 6 bar (90 psi) fit a pressure regulator
(LA512).
Connector
Collet
Blanking
transit plug
Pull plug
Depress
collet
Clean square
cut end
Tube 8mm OD
Simply push in tube to attach
Tube in secured position
Fitting Tubes
Feed Water Inlet Connection