ELGA
PURELAB Option-S 7/15 US
Operator Manual
PURELAB Option-S 7/15 US Version 1 08/14
Page 23
7.3 Cleaning the Inlet Strainer
The Inlet Strainer should be checked and cleaned every six months to
ensure that the strainer does not become clogged.
Step 1 - Remove the Inlet Strainer
1. SWITCH OFF electrical supply.
2. OPEN front door.
3. ISOLATE inlet water supply.
4. REMOVE the pre-treatment cartridge from its
clips and set aside to gain access to the inlet
strainer.
5. DEPRESS collars on both sides of strainer and
disconnect tubing.
6. REMOVE the inlet strainer from its position.
Step 2 - Dismantle the Inlet Strainer
1. HOLD inlet strainer over a sink or receptacle.
2. UNSCREW inlet strainer.
3. REMOVE mesh filter.
4. CHECK mesh filter for signs of wear or damage,
replace or clean as necessary.
Step 3 - Reassemble the Inlet Strainer
1. INSERT mesh filter into strainer, ENSURE it is
facing the correct direction.
2. SCREW up the inlet strainer.
Step 4 - Replace the Inlet Strainer
1. REPOSITION the inlet strainer.
2. REFIT tubes to inlet strainer, ENSURE it is facing
the correct direction.
3. REPOSITION the pre-treatment cartridge into its
support clips.
4. RE-ESTABLISH inlet water supply.
5. TURN on power.
7.4 Replacement of LC143 Reverse Osmosis Cartridge(s)
The reverse osmosis cartridge should be replaced if the permeate
water purity or flowrate is not adequate and does not meet predicted or
previous performance.
For information regarding the replacement of the LC143 Reverse
Osmosis Cartridge contact Customer Service.
WARNING!
ALL NEW RO CARTRIDGES CONTAIN
TRACES OF BACTERIOSTATIC SOLUTION.
THE RO CARTRIDGES WILL THEREFORE
REQUIRE RINSING PRIOR TO USE.
Inlet Strainer
Summary of Contents for ELGA PURELAB Option-S 15-US
Page 1: ...PURELAB Option S 7 15 US Operator Manual ...
Page 2: ......