Electrical
The unit can be connected universally to any electrical supply in
the range of
100 - 240V and 50 - 60Hz.
The mains lead is
supplied with a molded plug on one end and a molded
connector to the unit on the other. The unit should be connected
to an earth.
Drain
A semi rigid flexible connection to a sink or suitable drain
capable of handling at least 1.5 l/min is required. The drain point
should have a gravity fall below the level of the unit and any
connections direct to drain should have an air-break device
fitted.
Feed Water
The feed water should be of good quality and comply with
specifications provided.
This should enter the unit via an
8mm
(5/16") O/D
semi rigid tube, and should be in the temperature
range 1 to 35°C.
CAUTION!
Operating with feedwater temperatures
outside the range 1 to 35°C will cause
damage to the PURELAB Option-S US unit.
For pressurized feeds, the minimum direct inlet pressure is
4.0 bar (60 psi) and maximum inlet pressure is
6 bar (90 psi).
Higher feedwater pressures must be reduced using a pressure
regulator valve (Part No. LA512).
Reservoir feeds to the
PURELAB Option-S US
unit should be
positioned at the same height, or above the unit, to provide a
positive flooded inlet pressure.