Instructions for Use Evone
54
10 Technical Specifications
10.1 Evone Parameters
Parameter
Value / Limits
Units
Physical Parameters
Size L x W x H
399 x 320 x 287
mm
Weight
21
Kg
Dead Volume
42 mL (in configuration with Evone Breathing Tubing)
52 mL (in configuration with Evone Conventional
Tube Adapter)
(internal dead volume of Evone, i.e. breathing system)
mL
Electrical Parameters
Power Voltage
115 – 230
V
Frequency
50 / 60
Hz
Protection Class
Control Unit: Class I
-
Applied Parts
CO
2
Sensor, Breathing tubing, cartridge: Type BF
-
Power
120
VA
Minimum ventilation
time on full battery
~1
Hrs.
Fuse
T2AH 250V
-
Standard
Emissions class/group or Immunity test level
CISPR 11
Class: A / Group: 1
-
EN 61000-3-2
Class: A
-
EN 61000-3-3
PASS
-
IEC 61000-4-2
± 15 kV air
± 2 kV, ± 4 kV, ± 6 kV, ± 8 kV contact and air
-
IEC 61000-4-3
1 Radiated RF EM fields: 3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
RF wireless communications equipment:
Frequency:
Immunity test level:
385 MHz
27 V/m
450 MHz
28 V/m
710 MHz
9 V/m
745 MHz
9 V/m
780 MHz
9 V/m
810 MHz
28 V/m
870 MHz
28 V/m
-