Instructions for Use Evone
52
Alarm Text
Condition
Device Action
User Action
Priority
O
2
pressure low /
Set conditions
cannot be reached.
FiO
2
continues at
21%
Set conditions
cannot be reached
because of low O
2
supply pressure.
Continue
ventilation on
air supply. When
condition ceases
alarm is reset
and Evone
switches to both
gas supplies.
Check supply
pressure.
Consider to
connect to O
2
cylinder.
High
FiO
2
deviation / FiO
2
deviates from set
value
Mismatch in set
and measured FiO
2
percentage, for FCV
®
Mode > 5% point
and for Jet Mode
> 10% point.
Delay: 30 s
Alarm is reset
when condition
ceases.
Delay: 5 s
Check O
2
and air
supply. If alarm
persists, contact
manufacturer.
High
Gas supply pressure
too high / Gas supply
pressure > 8 bar.
Supply pressure
might damage
equipment.
Gas supply pressure
> 8 bar
Delay: 2 s
Move to safety
state.
Switch to
alternative gas
sources.
Consider to start
alternative
ventilation.
High
Battery low /
Battery level assures
maximal 30 minutes
of ventilation
Battery level not
sufficient for more
than 30 minutes of
ventilation.
Delay: 15 s
Alarm is reset
when condition
ceases.
Connect to mains
power.
Medium
Air pressure low /
Set conditions
cannot be reached.
FiO
2
continues at
100%
Set conditions
cannot be reached
because of low air
supply pressure.
Switch to 100%
O
2
after confir-
mation by user.
When condition
ceases alarm is
reset and Evone
switches to both
gas supplies.
Check supply
pressure.
Consider to
connect to air
cylinder.
Medium