Instructions for Use Evone
36
In the text in this instructions for use pressures are communicated in mbar or bar and EtCO
2
in mmHg. In table 6.2 and 6.3 calculated values into the other units that can be selected in
the ‘menu’ button are displayed. In Table 6.2 below, the ranges are given for all adjustable
ventilation parameters including the step size.
Parameter
Unit
Setting
Resolution
Default
Min
(Rounded)
Max
(Rounded)
(Rounded)
Peak
mbar
5
100
1
15
kPa
0.5
10.0
0.1
1.5
cmH
2
O
5
102
1
15
EEP
mbar
-10
30
1
5
kPa
-1.0
3.0
0.1
0.5
cmH
2
O
-10
31
1
5
I:E Ratio
-
1:1.0 – 1:2.5
0.1E
1 : 1.0
Inspiration Flow
L/min
2 – 20
1
12
FiO
2
Percentage
%
21 – 100
5 starting
from 25
50
Table 6.2 Setting ranges of parameters in FCV
®
mode