EMR
3
Setup and Operation Manual
C&C Mode Setup Categories
33
TEMPERATURE
Figure 35 illustrates Temperature Probe’s setup in C&C Mode.
Figure 35. Temperature Setup
Temperature Calibration
Figure 36 illustrates Temperature Probe Calibration setup choices.
Figure 36. Temperature Calibration Setup
FUEL SOURCE
METER CALIBRTN
CONFIGURE I/O
OEM MESSAGE
DELIVERY OPTIONS
REPORT FORMATS
RESTART
SECURITY CODE
LANGUAGE
DISPLAY SYNTAX
CELSIUS
FAHRENHEIT
Press ENTER to accept any changes in any column -
and/or continue pressing to return to TEMPERATURE
GO TO TEMPERATURE
CALIBRATION SETUP
+
-
+
-
+
Press the up/down
buttons to cycle
through remaining
C/C Mode categories
NEXT
ENTER
NEXT
TEMPERATURE
Press ENTER to accept the entry
and return to temperature units.
Press ENTER to return
to TEMPERATURE
Press ENTER after step
6 to accept the entry and
return to temperature units.
ENTER
CELSIUS
FAHRENHEIT
-
+
REFERENCE TEMP
CALIBRATE TEMP
-
+
NEXT
NEXT
ENTER
ENTER
Enter the REFERENCE TEMP
15˚C (60˚F). Press up/down arrows
to scroll through the numbers, and
NEXT to move from one digit of the
REFERENCE TEMP to the next.
NEXT
1. Fill the bottom (about 2-3 inches [100 mm])
of a one quart [1 litre] container with ice.
2. Center the temperature probe in the
container, then put ice all around it and on
top of it until the container is full.
3. Fill the container with water until it is level with the ice.
4. Verification should be performed in still air and
out ot direct sunlight.
5. Allow 2-5 minutes for the temperature measuring
devices to stabilize.
6. Select 0˚C [32˚F] for the CALIBRATE TEMP by pressing
the up/down buttons to scroll through the numbers and
NEXT to move from one digit of the CALIBRATE
TEMP to the next.