5. Installing the Equipment: Wall Mount
76
Smart Plus (PHT-35LHS) Installation Manual
5.6
Leveling the Equipment
T-shaped Hex
Wench
8 mm / 0.3”
Spirit Level
N/A
Ensure that the Spirit Level should rest only on the
locations indicated in the following figures to obtain the
accurate center.
1. Prepare the Spirit Level.
2. Turn the Rotating Unit clockwise so that the X-ray tube head faces the front as
shown in the figure.
3. Put the set screws into the four holes and turn them clockwise with the hex wrench
until they touch the aligning plate.
T-shaped Hex
Wench
8 mm / 0.3”
Set Screw
M10 x 20 – 4
pcs
(
Part No. 30)
4. Put the Spirit Level on the location as shown in the figure.
Summary of Contents for Smart Plus PHT-35LHS