2. Choosing an Installation Site
Smart Plus (PHT-35LHS) Installation Manual
15
En
glis
h
2.2
Specifications for Electrical Installation
These specifications are based on the
MEIGaN
(Medical electrical installation
guidance notes).
Consult the companion manual for further information:
Volume 3: Specification for
Electrical Installation
.
2.3
Electrical Requirement
This equipment must be connected to a grounded outlet to
fulfill the safety provisions specified in
IEC 60364: the 2
nd
edition (2006).
Both PC and equipment must use the same power line if
connected to an MPSO.
Whenever possible, use different power outlets for each
device. If a multiple portable socket outlet (MPSO) must be
used, ensure that the PC and equipment are connected to
the same MPSO.
Use a dedicated power outlet for the power cord. Failure to
do so may result in unstable system operations caused by
power fluctuations.
It is strongly recommended that you install an AVR. An AVR
(automatic voltage regulator) maintains a constant voltage
and allows for continuous operation in the event of power
fluctuation.
Item
Description
Power Supply Voltage
100 - 240 V ~
Frequency
50 / 60 Hz
Phase
Single
Power rating
2.0 kVA
Accuracy
Tube Voltage (kVp) ± 10 %, Tube Current (mA)
± 20 %, Exposure Time (s) ± (5 % + 50 ms)
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ±10 %.
Mode of operation: Continuous operation with intermittent loading - Needs
waiting time (at least 60 times the exposure time) before the next
Column operation time: Max. 2 min. On / 18 min. Off (Ratio 1:9)
Summary of Contents for Smart Plus PHT-35LHS
Page 1: ...Installation manual Model PHT 35LHS Version 1 29 English...
Page 2: ......
Page 148: ...Appendix A 2 Smart Plus PHT 35LHS Installation Manual Schematic Diagram Components Supplied...
Page 162: ...Appendix A 16 Smart Plus PHT 35LHS Installation Manual Intentionally left blank...
Page 164: ......
Page 165: ...www vatech com...