background image

4. Installing the Equipment: Base Stand (Optional) 

Smart Plus (PHT-35LHS) Installation Manual                                  

  51 

En

glis

Marking Points on the Wall

 

1.  Move the equipment to the installation site  as close as possible. 
2.  Adjust the distance between the wall and equipment by moving it slightly, so 

that  the wall bracket touches the wall. 

 

 

3.  Mark the anchor bolt locations on the wall. 

Marker 

N/A 

 

 

 

Summary of Contents for Smart Plus PHT-35LHS

Page 1: ...Installation manual Model PHT 35LHS Version 1 29 English...

Page 2: ......

Page 3: ...us is in steady improvement The information contained in this manual may be subject to change without notice justification or notification of the persons concerned VATECH Co Ltd manufacturer reserves...

Page 4: ...flat surface to maintain stability If the equipment is not stable property damage and personal injury may occur DO NOT push or pull the equipment Equipment should only be installed by an authorized te...

Page 5: ...erlock Switch 21 2 8 Installing the Emergency Stop Switch 21 3 Before Installing the System 22 3 1 Required Tools 22 3 2 Checking the ShockWatch and TiltWatch Indicators 24 3 3 Unpacking Boxes 25 3 4...

Page 6: ...Firewall Off 92 8 3 Setting up the Power Mangement Options 93 8 4 Turning off the User Account Control 95 8 5 Setting Folder Exclusions with Anti virus Software 96 9 Installing Software 97 9 1 Before...

Page 7: ...n by the user s manual PC Software has been properly used by the manufacturer s installation instructions and user manuals 1 2 Customer s Responsibility Site planning and preparation are the responsib...

Page 8: ...priate conditions or actions may cause Severe personal injury to the operator and patient Substantial property damage CAUTION Indicates that a potential hazard may exist which through inappropriate co...

Page 9: ...1 Introduction Smart Plus PHT 35LHS Installation Manual 3 English Never touch or hold the sensor or tube head areas while moving installing or operating the equipment...

Page 10: ...ile installing or operating the equipment DO NOT use the electric drill for tightening untightening screws Use it only for drilling a hole on the wall Grab the recommended holding areas only to transp...

Page 11: ...1 1 3rd edition 2005 Any equipment not approved by VATECH must comply with the applicable standards IEC 60950 1 2nd edition 2005 for IT equipment Ex PC and IEC 60601 1 3rd edition 2005 for medical ele...

Page 12: ...of this system shall comply with all local code requirements for electro medical systems IEC 60364 7 710 2002 It is strongly recommended that a UPS be installed at the same time as the equipment The...

Page 13: ...quipment IEC60601 2 7 and IEC60601 2 28 X ray Tube Heads IEC60364 7 710 2002 Local Code Requirements for Electro Medical System Installation IEC 60601 1 1 2005 regulation shall be met to their full ex...

Page 14: ...pplied Equipment IEC 60601 1 Degree of protection against leakage current and electric shock Class 1 equipment Label Radiation hazard Label Indicates the authorized representative in the European Comm...

Page 15: ...be collected separately Label Warns ESD hazard MCU board Board package Indicates that this equipment is classified as a CLASS 1 LASER PRODUCT by IEC 60825 1 ED 2 regulations Label Indicates that the...

Page 16: ...3 2008 IEC 60601 2 63 2012 21 CFR 1020 30 31 33 NEMA Standard publication PS 3 1 3 18 This is Class IIb equipment and obtained CE marking in April 2007 for regulations compliance in accordance with t...

Page 17: ...the operator should be as about the patient as possible This equipment should not be installed on thick carpets for stability reasons Anti static floor materials should be used around the equipment T...

Page 18: ...2 Choosing an Installation Site 12 Smart Plus PHT 35LHS Installation Manual With CEPH Unit optional 2805 9 mm L x 2 457 5 mm W or wider...

Page 19: ...2 Choosing an Installation Site Smart Plus PHT 35LHS Installation Manual 13 English Ceiling Height 2 437 5 mm H or higher...

Page 20: ...of the Entrance The door of the X ray room should have a clearance of more than 800 mm 31 5 wide Floor Area The floor of the X ray room must be stable and level for system balance The floor must be a...

Page 21: ...ensure that the PC and equipment are connected to the same MPSO Use a dedicated power outlet for the power cord Failure to do so may result in unstable system operations caused by power fluctuations...

Page 22: ...on panel with an over current circuit breaker rated for 20A must be used Maximally allowed deviation of the tube voltage tube current exposure time Tube Voltage kVp 10 Tube Current mA 20 Exposure Time...

Page 23: ...17 English 2 4 Environmental Specifications Item Description During Operation Temperature 10 35 Relative humidity 30 75 Atmospheric pressure 860 1060 hPa During Transport and Storage Temperature 10 60...

Page 24: ...are three options for installation depending on the configuration of the site Nevertheless the 2nd option is preferred Option No 1 The user operates the Exposure Switch from inside the X ray room Opti...

Page 25: ...art Plus PHT 35LHS Installation Manual 19 English Option No 3 The 3rd party Exposure Switch not VATECH s is used on the demand of the customers For this scenario see Appendix D Connecting the 3rd part...

Page 26: ...2 Choosing an Installation Site 20 Smart Plus PHT 35LHS Installation Manual 2 6 Installation Versions Base stand Type Wall mount Type...

Page 27: ...nergized The warning lamp and Door Interlock Switch are not included with the equipment The warning lamp and the Door Interlock Switch must be installed by a qualified technician 2 8 Installing the Em...

Page 28: ...Appendix F Installation Checklist 3 1 Required Tools The following tools are necessary to install the Smart Plus Item Figure Size Wrench Set 1 5 mm 10 mm 0 06 0 4 T shaped Hex Wrench 6 mm 10 mm 0 24 0...

Page 29: ...on Manual 23 English Item Figure Size Anti Static Glove N A Knife N A Tape Ruler for Wall Mount type 5 m Marker Pen thick tip for Wall Mount type N A Hammer for Wall Mount type N A Multimeter N A Hamm...

Page 30: ...ical shock or impact during transportation However the red indicator does not exactly mean that the equipment has been damaged These indicators are affixed only to the main box which contains the equi...

Page 31: ...I detector The digital CBCT system is based on a CMOS digital X ray detector The CMOS CT detector is used to capture 3D radiographic images of the head neck oral surgery implant and orthodontic treatm...

Page 32: ...eparate the top cover after removing the strapping bands 3 Lift a small distance remove the side covers 4 Remove the plastic wrap covering the box by using a cutter While one installer is removing a s...

Page 33: ...3 Before Installing the System Smart Plus PHT 35LHS Installation Manual 27 English PANO Box Configuration CEPH Box Configuration...

Page 34: ...tion 1 Pull out the upper and lower carrying handles from EPS Main Base 2 Assemble the upper carrying handle by using one wrench bolt Wrench Bolt M8 x 25 2 Sems 1 pcs Allen Wrench 6 mm 0 24 The Wrench...

Page 35: ...ch Bolt M8 x 30 2 pcs Allen Wrench 6 mm 0 24 The Wrench Bolt is attached to the Column unit Remove the bolts from the Column first 4 Remove the EPS Main Right and EPS Main Left as follows 1 Remove the...

Page 36: ...t the Main Unit with the EPS Main Base on a Transport Dolly and move it to the installation site When putting the Main Unit on the dolly make sure that the center of the mass indicator is aligned with...

Page 37: ...S Installation Manual 31 English 3 3 2 Box No 2 Base Unit Component Size mm inch Weight kg lbs Base 1035 L x 1025 W x 185 H 40 7 L x 40 3 W x 7 3 H 65 143 Removing the cover 1 Open the box cover and r...

Page 38: ...S Installation Manual 3 3 3 Box No 3 CEPH Unit Optional Component Size mm inch Weight kg lbs CEPH unit 1590 L x 820 W x 870 H 62 6 L x 32 3 W x 34 3 H 50 110 Removing the cover 1 Open the box cover an...

Page 39: ...3 Before Installing the System Smart Plus PHT 35LHS Installation Manual 33 English 3 4 Checking the Parts 3 4 1 Location Layout of the Parts and Accessories...

Page 40: ...No 2D Viewer Manual 1 For EasyDent i or EzDent Yes No 3D Viewer Manual 1 For Ez3D i or Ez3D Plus Yes No Installation USB 1 Yes No License Key 1 For Ez3D i and EzDent i Yes No 2 Exposure Switch 1 Yes N...

Page 41: ...available in some Asian countries Yes No Special A 1 Yes No Special B 1 Yes No 9 Chinrest Normal 1 Yes No 10 Left Blank Intentionally 11 Sanitary Vinyl Cover For Normal Bite 1 Yes No 12 Left Blank In...

Page 42: ...System Optic Cable 1 Yes No LAN Cable 1 Yes No IFG Card or FTG Card or LEG Card 1 Yes No MODEL Scan Jig 1 For the 3D MODEL Scan model Yes No 22 Column Bracket 1 Yes No 23 Wrench Bolt M10 x 25 w Sprin...

Page 43: ...No 30 Set Screw M10 x 20 4 Yes No 31 Protractor 1 Yes No Wall Plate 1 Yes No Terminal Block 3 Pole 1 Yes No 32 Middle Bracket 1 Optional Yes No Wrench Bolt M8 x 25 w Spring and Flat Washers 2 Yes No...

Page 44: ...Installation Manual Part No Items Specification Figure QTY Comments Confirmed OK Anchor Bolt 5 16 x 60 w Spring and Flat Washers 10 For Concrete Yes No 35 Wrench Bolt M8 x 25 w Spring and Flat Washers...

Page 45: ...cification Figure Qty Comments Location Confirmed OK Base Front Cover N A 1 N A On the upper side of the EPS Main Bottom Yes No Column Back Cover N A 1 N A On the upper side of the EPS Main Bottom Yes...

Page 46: ...e is a concrete floor go to section 4 4 Fixing the base Optional and do number 1 first after that turn back 4 1 Assembling the Base and Main Units 1 Remove the lower carrying handle Allen Wrench 6 mm...

Page 47: ...in the bolts as shown in the figure Hold the base unit to keep it from falling 4 Put the base unit to the column unit by engaging the base holes in the bolts as shown in the figure Allen Wrench 6 mm...

Page 48: ...lus PHT 35LHS Installation Manual 5 Tighten the two wrench bolts Allen Wrench 6 mm 0 24 6 Put the equipment in a vertical position slowly while holding the upper handle DO NOT damage the cables Before...

Page 49: ...ling the Equipment Base Stand Optional Smart Plus PHT 35LHS Installation Manual 43 English 7 Tighten the three wrench bolts to attach the base unit Allen Wrench 8 mm 0 31 Wrench Bolt M10 x 25 3 pcs Pa...

Page 50: ...ensor and tube head 1 Now it is assumed that the CEPH box has already been opened 2 Remove the plastic wrap covering the column unit and the tape for fixing the CEPH cables by using a cutter 3 Move an...

Page 51: ...uipment Base Stand Optional Smart Plus PHT 35LHS Installation Manual 45 English 4 Tighten the three wrench bolts to attach the base unit Allen Wrench 8 mm 0 31 Wrench Bolt M8 x 20 3 pcs Part No 25 Fla...

Page 52: ...e cables as shown in the figure When managing the fiber optic cable DO NOT bend pull or crush it Ensure that the caps of the fiber optic cable have been removed DO NOT touch the tip of the fiber optic...

Page 53: ...PHT 35LHS Installation Manual 47 English 6 Put the cables inside the CEPH Arm and assemble the CEPH arm cover by using two flat head screws Cross Head Screwdriver w Magnetic Tip 6 mm 0 24 Flat Head S...

Page 54: ...roperly installed Otherwise the equipment may be shaken during the scanning 1 Move the equipment to the installation site near the wall 2 Remove the lower carrying handle Allen Wrench 8 mm 0 31 One in...

Page 55: ...and Optional Smart Plus PHT 35LHS Installation Manual 49 English 5 Assemble the column bracket to the top ofthe column with four flat head screws Cross Head Screwdriver w Magnetic Tip 6 mm 0 24 Flat H...

Page 56: ...e the wall bracket Wall Bracket Part No 36 Wall Bracket for Wood Wall Optional Part No 37 2 Combine the column and wall brackets in the following manner with the 2 wrench bolts Allen Wrench 8 mm 0 31...

Page 57: ...nual 51 English Marking Points on the Wall 1 Move the equipment to the installation site as close as possible 2 Adjust the distance between the wall and equipment by moving it slightly so that the wal...

Page 58: ...wall holes of size 10 5 mm x 30 mm depth using the concrete hammer drill 2 Remove the debris and clean the holes using the dust pump 3 Using the hammer insert a Ficher strong anchor into the hole Fich...

Page 59: ...English Combining the Equipment with the Anchor Bolts 1 Place the equipment on the alignment plate while observing 4 Hex bolts are being appropriately inserted through each hole Hex Bolt M8 x 15 Spri...

Page 60: ...g and Flat Washers 2 pcs Part No 34 Hammer Drill L 200 mm 7 9 Hammer N A Ratchet Wrench Tips 3 mm 8 mm 0 12 0 3 1 Before installing the equipment put the base unit on the installation site and mark 2...

Page 61: ...PHT 35LHS Installation Manual 55 English 4 Remove nuts and washers put the anchor bolts into the holes 5 Secure the anchor bolts with the hammer 6 Place the base unit combined equipment in the proper...

Page 62: ...stalling the Equipment Base Stand Optional 56 Smart Plus PHT 35LHS Installation Manual Wooden Floor Wood Screws M12 x 70 2 pcs Part No 34 Hammer Drill L 200 mm 7 9 1 Secure the base unit using wood sc...

Page 63: ...4 5 Connecting the Cables to the Equipment 1 Connect the cables in the back of the column as shown in the figure Optic Cable Cable No H000014A Part No 21 Cable Tie Part No 20 Exposure Switch Part No...

Page 64: ...Plus PHT 35LHS Installation Manual 2 Assemble the Column Back Cover with four truss bolts Truss Bolt M4 x 8 2 pcs Part No 27 Cross Head Screwdriver w Magnetic Tip 6 mm 0 24 3 Ensure the cables go thr...

Page 65: ...t Plus PHT 35LHS Installation Manual 59 English 4 6 Removing the Transportation Safety Bolts 1 Remove the two column fixing bolts including the tags Allen Wrench 6 mm 0 24 2 Remove the semi clear tape...

Page 66: ...nual 3 Remove the Vertical Frame Cover 4 Remove four Safety Bolts and two Safety Brackets including the tags Allen Wrench 6 mm 0 24 Be sure not to damage the cables when removing the bolts The Safety...

Page 67: ...irit Level N A Ensure that the Spirit Level should rest only on the locations indicated in the following figures to obtain the accurate center 1 Prepare the Spirit Level 2 Turn the Rotating Unit clock...

Page 68: ...2 Smart Plus PHT 35LHS Installation Manual Leveling Right and Left 1 Place the Spirit Level as shown in the figure 2 Adjust the base until the bubble on the Spirit Level centers in the middle by turni...

Page 69: ...own in the following figure 2 Adjust the screws until the bubble of Spirit Level centers level by turning the front and back screws clockwise or vice versa 3 When the leveling is completed make sure t...

Page 70: ...he Equipment You are advised to plan and study the installation environment carefully in advance before proceeding since the installation involves drilling the wall and floor Pre installation planning...

Page 71: ...LHS Installation Manual 65 English Marking Points on the Floor 1 Put the aligning plate on the floor near the wall where the equipment should be installed as shown in the figure Alignment Plate Part N...

Page 72: ...Drill the floor holes size 10 5 x 30 mm by using the concrete hammer drill 3 Remove the debris and clean the holes withthe dust pump 4 Put the anchor bolts into the holes and attach them with the ham...

Page 73: ...nchor Bolts 1 Put the aligning plate on the floor and ensure that all anchored bolts come through the holes on the plate 2 Put the equipment in a vertical position slowly while holding the upper handl...

Page 74: ...the anchored bolts 4 Put the washers and nuts into the six anchored bolts on the floor and tighten the nuts loosely Make sure that you put the fasteners in the sequence as shown in the figure DO NOT...

Page 75: ...5 3 Installing the Wall and Column Brackets Assembling the Column Bracket 1 Move the equipment to the installation site near the wall 2 Remove the lower carrying handle Allen Wrench 6 mm 0 24 One inst...

Page 76: ...Wall Mount 70 Smart Plus PHT 35LHS Installation Manual 5 Assemble the column bracket to the top ofthe column with four flat head screws Cross Head Screwdriver w Magnetic Tip 6 mm 0 24 Flat Head Screw...

Page 77: ...e the wall bracket Wall Bracket Part No 36 Wall Bracket for Wood Wall Optional Part No 37 2 Combine the column and wall brackets in the following manner with the 2 wrench bolts Allen Wrench 6 mm 0 24...

Page 78: ...Manual Marking Points on the Wall 1 Move the equipment to the installation site as close as possible 2 Adjust the distance between the wall and equipment by moving it slightly so that the wall bracke...

Page 79: ...all holes of size 10 5 mm x 30 mm depth using the concrete hammer drill 2 Remove the debris and clean the holes using the dust pump 3 Using the hammer insert a Ficher strong anchor into the hole Fiche...

Page 80: ...Combining the Equipment with the Anchor Bolts 1 Place the equipment on the alignment plate while observing 4 Hex bolts are being appropriately inserted through each hole Hex Bolt M8 x 15 Spring washer...

Page 81: ...S Installation Manual 75 English 5 4 Connecting the Cables to the Equipment Please refer to section 4 5 Connecting the Cables to the Equipment 5 5 Removing the Transportation Safety Bolts Please refer...

Page 82: ...the following figures to obtain the accurate center 1 Prepare the Spirit Level 2 Turn the Rotating Unit clockwise so that the X ray tube head faces the front as shown in the figure 3 Put the set scre...

Page 83: ...ng the Equipment Wall Mount Smart Plus PHT 35LHS Installation Manual 77 English 5 Turn each set screw clockwise or counterclockwise to make the equipment level while another person monitors the level...

Page 84: ...Equipment Wall Mount 78 Smart Plus PHT 35LHS Installation Manual 5 7 Tightening the Bolts 1 Tighten the joint bracketbolts Allen Wrench 6 mm 0 24 Monkey Wrench N A 2 Tighten the nuts in the anchored b...

Page 85: ...nual 79 English 6 Completing Miscellaneous Works 6 1 Assembling Various Covers Base Cover Optional Truss Bolt M5 x 8 3 pcs Part No 26 Base Cap 3 pcs Part No 17 1 Assemble the base cover and fix it wit...

Page 86: ...you do not use the base during the installation remove 2 pcs of Truss bolts while installing the front cover as below Truss Bolt M4 x 8 2 pcs Part No 27 Silicon Cap 2 pcs Part No 16 1 Assemble the ba...

Page 87: ...ts and the Chinrest Assembling the Temple Supports and the Bite Block should be done after acquiring a test image is completed 1 Insert the normal Chinrest and the normal Bite Block into the unit Chin...

Page 88: ...o 33 Truss Bolt M4 x 10 2 pcs Part No 33 UP DOWN Front Cover 1 Assemble the UP DOWN Switch Holder on the bottom of the Vertical Frame with two Truss Bolts M4 x 10 2 Connect the UP DOWN switch to the u...

Page 89: ...Installation Manual 83 English 7 Setting up PC 7 1 Direct Connection Diagram Fiber Optic Cable Used to transfer image data to thePC Warning System Panel Used to provide a visible indicator Light on wh...

Page 90: ...egulator to keep the line voltage stable The PC system provided with the Smart Plus undergoes the rigorous test for software compatibility before shipping Therefore any later changes to the hardware a...

Page 91: ...ot enabled Windows 10 is not protected from malware and virus Disabling Windows Defender 1 Open the Start screen type Windows Defender in the search box 2 Click the Windows Defender icon to start Wind...

Page 92: ...7 Setting up PC 86 Smart Plus PHT 35LHS Installation Manual 5 On Select Folder window type C VcaptureSW in the folder field and click Exclude this folder...

Page 93: ...power cord from the main outlet or PC Disregard this section in case the PC system is supplied with the equipment The peripherals have already been installed inside the PC Whenever managing the fiber...

Page 94: ...3 Insert the frame grabber Part No 21 carefully into that PCIe2 x 4 slot and lock it IFG Card or FTG Card or LEG Board Double check the locking status between the board and its holder after the board...

Page 95: ...ition arise from the bad cable or its bad contact condition Connect the regular cables for PC keyboard mouse and video in advance The following figures and descriptions are based on the PC model Z4 fr...

Page 96: ...7 Setting up PC 90 Smart Plus PHT 35LHS Installation Manual 4 Confirm the result after connections are as same as below The illustrations may differ from the actual product...

Page 97: ...r USB imputation of images should be installed on the formatted PC where no other program s except the operating system OS is present Before InstallShield installation ensure that the video card drive...

Page 98: ...or this reason it is required to disable the Windows Firewall by using the following procedures 1 Open the Start screen type Windows Firewall in the search box 2 Click the Windows Firewall icon to sta...

Page 99: ...s To avoid disruptive and abnormal operation while acquiring an image it is required to reconfigure some parameters on the Windows operating system Disable the Screen Saver 1 Open the Start screen typ...

Page 100: ...ual Selecting the Power Options Monitor and System 1 Open the Start screen type Power Options in the search box 2 Click Choose when to turn off the display 3 Select Never for both Turn off the display...

Page 101: ...LHS Installation Manual 95 English 8 4 Turning off the User Account Control 1 Open the Start screen type User Account Control in the search box 2 Disable the UAC by moving the slider bar down to the b...

Page 102: ...scans performed by the anti virus software For Smart Plus the following folders and files inside for relevant software should be excluded from the virus scan Path Software C ProgramFiles Vatech Ez3D i...

Page 103: ...ed Downloading irrelevant programs to the same PC may cause conflicts between programs and result in equipment failures The image viewer program such as EzDent i EasyDent or the one from the 3rd party...

Page 104: ...l 9 2 Software Installation Flow 9 3 Installing Image Viewer Program One of the image viewer programs among the EzDent i EasyDent or 3rd party program must be installed at this time For the details on...

Page 105: ...lease check the serial number txt e g 047 011752 txt file 1 Turn on the PC and the equipment if they are not yet 2 Insert the USB drive into the USB connector and then perform a virus scan for the PC...

Page 106: ...ecked as below If the CEPH feature is not included in the equipment uncheck the CEPH option Moreover click Next 7 Confirm that the WidePano is selected in the CBCT sensor type selection window and cli...

Page 107: ...on window and clickNext 10 Select the default port number COM2 Select the port number COM2 The same COM port number should be used between the equipment and PC 11 Select the language and click Next 12...

Page 108: ...Next to continue When EasyDent is installed select EasyDent when EzDent i is installed selectSDK 14 Check the options according to the product specifications When the AutoSave is checked the image da...

Page 109: ...PHT 35LHS Installation Manual 103 English 16 You can check the information entered so far with the Ready to Install window If necessary you can modify it by clicking Back 17 When the information is co...

Page 110: ...nual Installing the DICOM Viewer 1 After completing basic settings the DICOM Viewer installation will be started Click Next in the Welcome window 2 Select I accept the terms of the license agreement a...

Page 111: ...9 Installing Software Smart Plus PHT 35LHS Installation Manual 105 English 4 From the next screen click Install 5 Click Finish to exit the wizard...

Page 112: ...lus PHT 35LHS Installation Manual Installing the DirectX 1 After completing the DICOM viewer installation DirectX installation will be started Select I accept the agreement in the Welcome window 2 Cli...

Page 113: ...9 Installing Software Smart Plus PHT 35LHS Installation Manual 107 English 3 Click Finish to exit the wizard...

Page 114: ...Inform Grabber 1 After completing the Direct X installation Virtual Serial installation will be started Click Install in the Welcome window 2 During installation VA_4x Cap 1 1 installation wizard wil...

Page 115: ...sual C 2010 x64 Redistributable installation or maintenance window will appear Select one of the options and click Next to continue the process 6 Click Finish to exit the Visual C 2010 x64 Redistribut...

Page 116: ...fter completing the Direct X installation Grabber Driver Setup will be started Select the language in the installer language window and press the OK button 2 When the Grabber Driver Setup window appea...

Page 117: ...appear and disappear and the Completing Grabber Driver Setup window will appear Choose the Reboot now or I want to reboot later and click the button manually If an error occurs during installation it...

Page 118: ...the installation program will be started Select the language in the installer language select window and press the OK button 2 Press the V Grabber s Install button 3 The installation is completed and...

Page 119: ...nstall_Log txt on the desktop 2 Open it to check the file You can find out that all components are installed 3 Go to section 9 5 Setting up the User specific Information Uninstalling Software 1 Open t...

Page 120: ...m EasyDent or EzDent i SDK installed on the PC EasyDent Refer to Section 9 5 1 when the EasyDent is installed EzDent i Refer to Section 9 5 2 when the EzDent i is installed 9 5 1 When the EasyDent is...

Page 121: ...n Manual 115 English Interfacing the EasyDent with Imaging Program One time Linking 1 On the main screen of the EasyDent click Help Configuration 2 Click on the Program tab 3 Check Select capture mode...

Page 122: ...Installing Software 116 Smart Plus PHT 35LHS Installation Manual 4 Click on the button for the Path and select the Console SW s executable file 5 The Path is displayed as shown in the following figur...

Page 123: ...ware Smart Plus PHT 35LHS Installation Manual 117 English 6 Check Select capture mode for panorama and select Etc from the drop down list 7 Click on the button for the Path and select the Console SW s...

Page 124: ...35LHS Installation Manual 8 The Path is displayed as shown in the following figure 9 Click Apply 10 When the following message appears click OK to restart the program 11 Re run the EasyDent program an...

Page 125: ...nying EasyDent Manual 1 Click the Patient icon on the upper left corner of EasyDent s main GUI window 2 Then the following dialog box will be displayed Enter the required patient information Chart No...

Page 126: ...imaging program the registration window will be displayed 1 Click Next to proceed 2 Select I agree to the Vatech Privacy Policy and click Next 3 Enter the required information and click Next For the...

Page 127: ...ole PC connected to the Internet 4 Click Finish For the Console PC NOT connected to the Internet 1 Click Download file to download and save the file HTML type into a memory device 2 Execute the file i...

Page 128: ...re 122 Smart Plus PHT 35LHS Installation Manual 4 Download auth Acer file 5 Copy the downloaded auth cert file to the console PC 6 Click the Upload Verification file to upload auth cert file 7 Click N...

Page 129: ...rner of EasyDent s main window to open the imaging program 2 The main screen of the selected mode is displayed From the main screen you can configure the imaging parameter settings before acquiring an...

Page 130: ...nstallation Manual 9 5 2 When the EzDent i is installed This program must be run in administrator mode Running the Image Viewer 1 Run the image viewer On your desktop double click the EzDent i icon Th...

Page 131: ...Smart Plus PHT 35LHS Installation Manual 125 English Interfacing the EzDent i with Imaging Program One time Linking 1 On the main screen of the EzDent i click EzDent i Setting 2 Click Acquisition Con...

Page 132: ...3 Make sure that the console program settings are as follows Console Program Path C VCaptureSW exe VCaptureSW exe Capture Message Smart Plus Captured Patient Information File Path C VCaptureSW Patien...

Page 133: ...ect refer to the accompanying EzDent i Manual 1 Click on the Add Patient button from the PATIENT tab 2 Enter the required patient information Chart Number First Name and Last Name are required fields...

Page 134: ...differently depending on the equipment s capacity for acquiring images 2 Select the imaging mode Then the main GUI in the selected imaging mode appears The error code E033 indicates that the equipment...

Page 135: ...tered following the user requirements 1 From the main GUI window click the settings icon in the upper right corner 2 Log in the tap will be open by default On the right side of the screen select Engin...

Page 136: ...the following figure 6 In the DAP check that the default DAP Level and DAP Unit are displayed as shown in the following figure 7 In the Language Option the default language setting is English If nece...

Page 137: ...formation Setting configure the Link Type File Path and File Name as shown in the following figure 10 In the Link Information Setting configure the Link Type and File Extension as below Fields When Ea...

Page 138: ...132 Smart Plus PHT 35LHS Installation Manual 11 Click on the Default tab and configure the user defined parameters The default features can be modified according to the user s requirements 11 Click Sa...

Page 139: ...ed the equipment will not be operated even after being turned on 1 Click on the General tab on the Control Panel 2 In the Networking option enable the Serial checkbox and set Serial port and Baud Rate...

Page 140: ...PKEN_0000 to exit the packing mode Now note that the equipment is out of the packing mode Use the command PKEN_0001 to re enter the packing mode 5 Click Exit and terminate the Control Panel 6 Importan...

Page 141: ...PANO Music Different for each mode 0001 CT PANO Music Same for each mode 0002 Default CT PANO Beep Same for each mode 1 Send the command following the command specification as specified in the table...

Page 142: ...9 Installing Software 136 Smart Plus PHT 35LHS Installation Manual Finalizing the Parameter Settings 1 Click Exit Close and terminate the Control Panel 2 Important Exit the imaging program main GUI...

Page 143: ...Smart Plus PHT 35LHS Installation Manual 137 English 10 Technical Specifications 10 1 Mechanical Specifications Dimensions unit mm With CEPH unit Without CEPH unit Top view Side view Wall bracket dime...

Page 144: ...t Without CEPH unit 1144 3 L x 1557 5 W x 2306 H mm without Base 45 1 L x 61 4 W x 90 8 H inch without Base 1144 3 L x 1557 5 W x 2337 5 H mm with Base 45 1 L x 61 4 W x 92 1 H inch without Base With...

Page 145: ...uracy Tube Voltage kVp 10 Tube Current mA 20 Exposure Time s 5 50 ms The input line voltage depends on the local electrical distribution system Allowable input voltage fluctuation requirement 10 Mode...

Page 146: ...allation Manual 10 3 Environmental Specifications Item Description During Operation Temperature 10 35 Relative humidity 30 75 Atmospheric pressure 860 1060 hPa Transport and Storage Temperature 10 60...

Page 147: ...agreed with the local radiation protection advisor RPA customer 3 A fluorescent lamp shall not be used in the X rays on sign 4 The customer shall be responsible for the proper installations for the wa...

Page 148: ...Appendix A 2 Smart Plus PHT 35LHS Installation Manual Schematic Diagram Components Supplied...

Page 149: ...m 198 Warning lamp 1 m 40 Power source cable 1 m 40 1 Prepare the Warning System Panel Part No 28 2 Install the Warning System Panel at the proper height after taking each cable length into account 3...

Page 150: ...the emergency case but cannot be pushed by mistake The switch shall be a type of mistake proof The switch is not supplied The switch shall be installed at a height of 1 2 to 1 5 meters 47 to 60 1 The...

Page 151: ...t This section explains how to limit the column height within the permissible range 1 Measure the ceiling height in the X ray shield room Hceiling Removing the column covers 2 Remove two Fixing Bolts...

Page 152: ...he top of the equipment when the column is fully extended d Hceiling HMax Hceiling 2306 mm Example height The height of the equipment without Base Ex If Hceiling is 2200 mm Hscrew height value is calc...

Page 153: ...m the default current position to the new one 1 Loosen two bolts halfway important DO NOT unscrew completely the bolt If not it could drop into the column and may cause big trouble to retrieve it out...

Page 154: ...connect the third party Exposure Switch with the equipment from VATECH How to 1 Cut the Exposure Switch cable provided with the equipment 2 According to the following schematic diagram rewire the cabl...

Page 155: ...C BIOS Settings HP PC BIOS Setup PC Model HP Z4 PC BIOS default Main Menu Sub1 Menu Sub2 Menu Setup Value Advanced Power Options Runtime Power Management Disable Advanced Power Options Idle Power Savi...

Page 156: ...anual F Installation Checklist 1 General Information Customer Information about the Equipment Purchaser Name of Clinic or Hospital Address Phone E Mail Website Dealer Information about the Equipment S...

Page 157: ...stallation Date of installation Model Serial No 3 System Delivery to Site Yes No Did you review and identify the delivery route and method for equipment in advance Is the freight elevator available Is...

Page 158: ...Does the electrical input condition to the installationsite meet the MEIGaN requirements Is the local Network IP address of clinic 192 168 33 xx Is a compressor or air conditioner suction located righ...

Page 159: ...ny suspicious holes or scratches on the box after opening 5 While Installing Equipment Yes No Are installers careful with any sensitive parts while carrying equipment Did the installers make sure that...

Page 160: ...Do the equipment Up Down switch works properly 7 Software Compatibility Yes No Anti virus software installed A firewall installed If yes indicate software or hardware Type Is the 3rd party software i...

Page 161: ...Configuration Yes No Is a network configured with 1 Gbit s of CAT5 Is the equipment connected to thenetwork Is the network installation company identified What is the TCP IP addressassigned Address W...

Page 162: ...Appendix A 16 Smart Plus PHT 35LHS Installation Manual Intentionally left blank...

Page 163: ...n permission of the manufacturer We reserve the right to make any alterations that may be required due to technical improvement For the most current information contact your VATECH representative Tel...

Page 164: ......

Page 165: ...www vatech com...

Reviews: